3
Set up the Trigger menu.
a
Press the Trigger key.
b
Select the Mode/Arm field, then select Edge.
c
Select the Source field, then select C1.
d
Move the cursor to the Level field. Set the trigger level to 0 by typing 0 in the
front-panel keyboard, then pressing Select.
4
Turn off the voltage and time markers.
a
Press the Marker key.
b
Move the cursor to the V Markers field and press Select. The Select key should toggle
the marker to Off.
c
Move the cursor to the T Markers field and press Select. At the pop up menu, select
Off.
Testing Performance
To test the bandwidth (oscilloscope)
3–74