![Agilent Technologies 1670G Series Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830076.webp)
7
Test the next setup/hold combination.
a
In the logic analyzer Format menu, select Master Clock.
b
Turn off and disconnect the clock just tested.
c
Repeat steps 1 through 10 for the next setup/hold combination listed in step 1 on
page 3–48, until all listed setup/hold combinations have been tested.
When aligning the data and clock waveforms using the oscilloscope, align the waveforms
according to the setup time of the setup/hold combination being tested, +0.0 ps or
−
100 ps.
Test the next channels
•
Connect the next combination of data channels and clock channels, then test them.
Start on page 3–44, "Connect the logic analyzer," connect the next combination, then continue
through the complete test.
Testing Performance
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
3–52
Содержание 1670G Series
Страница 20: ...1 12...
Страница 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Страница 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Страница 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Страница 186: ...Exploded View Exploded view of the Agilent 1670G series logic analyzer Replaceable Parts Exploded View 7 4...
Страница 197: ...The Logic Acquisition Board Logic Acquisition Board Block Diagram Theory of Operation The Logic Acquisition Board 8 7...
Страница 200: ...The Oscilloscope Board Oscilloscope Board Block Diagram Theory of Operation The Oscilloscope Board 8 10...
Страница 201: ...Theory of Operation The Oscilloscope Board 8 11...