![Agilent Technologies 1670G Series Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830081.webp)
4
Enable the pulse generator channel 1 output (with the LED off).
5
Set up the Waveform menu.
a
Press Run to fill the acquisition memory so the markers can be configured.
b
Press the Waveform key.
c
Move the cursor to the sec/Div field, then use the knob to dial in 100 ns.
d
Select the Markers Off field, then select Pattern.
e
Select the Specify Patterns field. Select X entering 1 and O entering 1.
f
Select Done to exit the Specify Patterns menu.
g
Move the cursor to the X-pat field. Type 1, then press Done.
h
Move the cursor to the O-pat field. Type 16384, then press Done.
i
Select O-pat from Trigger field, and then select from X marker.
Testing Performance
To test the time interval accuracy (logic analyzer)
3–57
Содержание 1670G Series
Страница 20: ...1 12...
Страница 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Страница 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Страница 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Страница 186: ...Exploded View Exploded view of the Agilent 1670G series logic analyzer Replaceable Parts Exploded View 7 4...
Страница 197: ...The Logic Acquisition Board Logic Acquisition Board Block Diagram Theory of Operation The Logic Acquisition Board 8 7...
Страница 200: ...The Oscilloscope Board Oscilloscope Board Block Diagram Theory of Operation The Oscilloscope Board 8 10...
Страница 201: ...Theory of Operation The Oscilloscope Board 8 11...