![Agilent Technologies 1670G Series Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830040.webp)
Test the next pod
1
Using the 17-by-2 test connector and probe tip assembly, connect the data and clock
channels of the next pod to the output of the function generator until all pods have
been tested.
2
Unassign the channels just tested.
a
Press the Format key.
b
Move the cursor over the channel assignment field for the pod tested. Press Clear
Entry until all channels show unassigned (all ’.’).
c
Move the cursor over the Data on Clocks channel assignment field. Press Clear Entry
until all channels show unassigned (all ’.’).
3
Start with "Set up the logic analyzer" on page
3−
10, substituting the next pod to be
tested for pod 1.
Testing Performance
To test the threshold accuracy (logic analyzer)
3–16
Содержание 1670G Series
Страница 20: ...1 12...
Страница 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Страница 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Страница 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Страница 186: ...Exploded View Exploded view of the Agilent 1670G series logic analyzer Replaceable Parts Exploded View 7 4...
Страница 197: ...The Logic Acquisition Board Logic Acquisition Board Block Diagram Theory of Operation The Logic Acquisition Board 8 7...
Страница 200: ...The Oscilloscope Board Oscilloscope Board Block Diagram Theory of Operation The Oscilloscope Board 8 10...
Страница 201: ...Theory of Operation The Oscilloscope Board 8 11...