3
Set up the Trigger menu.
a
Press the Trigger key.
b
Select the Mode/Arm field, then select Edge.
c
Select the Source field and set it to C1.
d
Move the cursor to the Level field. Set the trigger level to 0 by typing 0 in the
front-panel keyboard, then pressing Select.
e
Select the Slope field and set it to Positive.
4
Set up the Markers menu.
a
Press the Marker key.
b
Move the cursor to the T Markers field and press Select. At the pop up menu, select
Auto.
c
Select the X marker, and set it on C1 at Level 50%, Slope Positive, Occur 1 (use the
RPG knob for Occur).
d
Select the O marker, and set it on C1 at Level 50%, Slope Positive, Occur 2 (use the
RPG knob for Occur).
e
Select the Statistics field and set it to On.
f
Select Done.
Testing Performance
To test the time measurement accuracy (oscilloscope)
3–79