![ACOEM Met One Instruments GAS-1060 Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/acoem/met-one-instruments-gas-1060/met-one-instruments-gas-1060_user-manual_459535105.webp)
Service
Page 105
Monthly
Perform precision check (pre-check) prior to
commencing any service tasks or making any changes
to the system as found in its current state. This task is
necessary to validate any previously captured data.
Check particulate filter, replace if full/dirty
Perform a pressure check
Check sample inlet system for moisture or foreign
materials. Clean if necessary
Check fan filter and clean if necessary
Check event log
Check date and time is correct
Check instrument status light
External vacuum pump check (vacuum source)
Or
Internal pump check
Leak check
Perform a manual background, record background
reading ensure it is less than 600 Mm-1. Observe the
trend and if the reading will increase to > 650 Mm-1
within the next 6 months perform a background
scrubber DAU check.
Perform NO
2
span calibration
Perform precision check (post-check) once all service
tasks have been completed. This task is necessary to
establish a valid start point to begin capturing new
data.
3 Monthly
Replace sample DFU (depending on site conditions) * 6.3.4
6 Monthly
Perform multipoint precision check
Calibrate analog outputs (only if used)
Yearly
Replace Background Scrubber DAU (depending on
site conditions) *
Flow check, replace or clean orifice (only if necessary)
* Suggested intervals for maintenance procedure are a guide only and may vary with sampling intensity and/or
environmental conditions. Please refer to your local regulatory standard for your personalised maintenance
schedule.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...