Serinus 60 User Manual 3.0  

 

Page 56 

 

 

3.4.26

 

Digital Input Menu 

Main Menu 

 Communications Menu 

 Digital Input Menu 

This menu is used to remotely trigger zero and span calibrations. This is done by assigning the eight 
digital inputs with one of the following commands. 

DI N (Pin X) 

Associates an action with a digital input. There are eight 
available digital inputs (the pin numbers are for the 25-pin 
connector). Each one can have one of the following associated 
actions, triggered when the corresponding digital input goes 
to the 

Active

 state: 

 

Disabled:

 No action (this digital input does nothing) the pin 

is forced to Active Low. 

 

Do Span:

 Used to perform a span check When activated the 

instrument sets the 

Cal. Mode

 to 

Span (

refer to Section 

3.4.10.1). 

 

Do Zero:

 Used to perform a zero When activated the 

instrument sets the 

Cal. Mode

 to 

Zero 

(refer to Section 

3.4.10.1)

Active 

[DO Span, Do Zero] 

Each pin can be set to be active 

High

 or 

Low

. Active High 

means that the event will be triggered when the line is pulled 
to 5V. Active Low means that the event will be triggered when 
the line is pulled to 0V. 

 

Example 

Here is a typical configuration between an instrument and either a data logger or calibrator (master 
device): 

1.

 

Set the jumper JP1 to 5V position (refer to Section 4.4.3). 

2.

 

Connect one of the master devices digital output signals to pin 18 and the ground signal to pin 5 
of the instrument’s analog/digital 25 pin female connector (refer to Figure 18)

3.

 

Program master device to output 0 volts to pin 18 when a span is desired. 

4.

 

In the instrument’s 

Digital Input Menu

 assign 

DI 0 

 Do Span

.  

5.

 

Set the 

Active

 state to 

Low

 

6.

 

The same procedure can be followed to also activate zero calibrations. Pin 6 of the instrument’s 
analog/digital 25 pin female connector can be connected to one of the other master devices 
digital outputs and the instrument can be set so 

DI 1

 is assigned to 

Do Zero

3.4.27

 

Digital Output Menu 

Main Menu 

 Communications Menu 

 Digital Output Menu 

This allows the instrument to trigger external alarms in response to certain events. 

Содержание Met One Instruments GAS-1060

Страница 1: ...USER MANUAL metone com GAS 1060 Direct Nitrogen Dioxide Analyzer US EPA Designated Equivalent Method Version 3 0...

Страница 2: ...One Instruments have chosen to use the original Ecotech Serinus operating manual Although the name of the Met One instrument is different the analyzer itself has remained unchanged and meets the stric...

Страница 3: ...2 6 Communications 20 1 2 7 Physical Dimensions 21 1 3 Nomenclature 21 1 4 Background Theory 22 1 4 1 Measurement Theory 23 1 5 Instrument Description 24 1 5 1 Calibration Valve Manifold 24 1 5 2 Bac...

Страница 4: ...nu 49 3 4 15 Internal Pump Menu Option 50 3 4 16 Valve Menu 50 3 4 17 Tests Menu 51 3 4 18 Digital Input Test Menu 51 3 4 19 Digital Output Test Menu 52 3 4 20 Calculation Factors Menu 52 3 4 21 Commu...

Страница 5: ...Calibration 89 5 2 2 Ambient Point Pressure Calibration 90 5 3 Pressure Calibration Internal Pump Option Only 91 5 4 Zero Calibration 91 5 4 1 Calibration Port 92 5 4 2 Sample Port 92 5 4 3 Internal Z...

Страница 6: ...Internal Pump PN E020124 125 8 2 1 Hardware Setup 125 8 2 2 Pressure Calibration 126 8 2 3 Pump control 126 8 3 Rack Mount Kit PN E020116 126 8 4 High Pressure Zero Span Valves 130 9 Parts List 131 9...

Страница 7: ...tion Prompt 72 Figure 23 Successful Driver Installation 72 Figure 24 Airodis Workspace Manager 74 Figure 25 Adding a New Station 74 Figure 26 New Station Connection 75 Figure 27 Station Configuration...

Страница 8: ...9 Leak Check Troubleshooting Flow Chart 117 Figure 60 Zero Flow Fault Troubleshooting Flow Chart 118 Figure 61 Noisy or Unstable Readings Troubleshooting Flow Chart 119 Figure 62 Flow Block Temp failu...

Страница 9: ...Annual Maintenance Kit PN E020203 01 131 Table 11 Consumables 132 Table 12 Spare Parts List 132 Table 13 Packet Format 141 Table 14 Example Primary Gas Request 141 Table 15 Example Primary Gas Respon...

Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...

Страница 11: ...till unsure or unclear on any part of the Serinus 60 please do not hesitate to contact Ecotech or your local Ecotech distributor Please help the environment and recycle the pages of this manual when y...

Страница 12: ...Always disconnect the power source before removing or replacing any components Warranty This product has been manufactured in an ISO 9001 ISO 14001 facility with care and attention to quality The prod...

Страница 13: ...f calling from within Australia Please dial 61 3 9730 7800 if calling from outside of Australia Please contact Ecotech and obtain a Return Material Authorisation RMA number before sending any equipmen...

Страница 14: ...munity IEC 61000 4 5 Surge immunity IEC 61000 4 6 Conducted RF immunity IEC 61000 4 11 Voltage dips and interruption immunity Electromagnetic Compatibility EN 61326 1 CISPR 11 Radiated RF emission mea...

Страница 15: ...ocuments for your records Please refer to the instrument name model number serial number sales order number and your purchase order number on all claims You should also Contact you freight forwarder f...

Страница 16: ...d Symbols on Ecotech Equipment Protective conductor terminal IEC 60417 5017 Alternating current IEC 60417 5032 Caution hot surface IEC 60417 5041 Caution risk of danger Refer to accompanying documents...

Страница 17: ...ation Any information that cannot be found within this manual can be obtained by contacting Ecotech This manual uses cross reference links extensively throughout this manual The hot keys below will gr...

Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...

Страница 19: ...rs the detector response and many other parameters The concentrations are automatically corrected for gas temperature and pressure changes and when using gravimetric units e g g m is referenced to 0 C...

Страница 20: ...up 65 VA after warm up 1 2 5 Operating Conditions Ambient Temperature Range 0 C to 40 C 32 F to 104 F Sample Pressure Dependence 5 change in pressure produces less than a 1 change in reading Maximum a...

Страница 21: ...ed 1 2 7 Physical Dimensions Case Dimensions Rack length front to rear 597 mm 23 5 Total length with latch release 638 mm 25 1 Chassis width 418 mm 16 5 Front panel width 429 mm 16 9 Chassis height 16...

Страница 22: ...al diameter of tubing and OD is the outer diameter Multidrop A configuration of multiple instruments connected via the same RS232 cable Bootloader A program that checks whether the current firmware is...

Страница 23: ...with a narrow band wavelength of 450 nm is used as a photon source for the cavity absorption process The LED is pulsed into the optical cavity and the photon leakage as a result of the optical cavity...

Страница 24: ...Internal Components Diagram 1 5 1 Calibration Valve Manifold The calibration valve manifold switches between sample calibration and background gas 1 5 2 Background Scrubber DAU The background scrubbe...

Страница 25: ...source and splits the frequency signal from the Mixer Preamp PCA into the 4 separate phases for the Main Controller PCA Various power supplies and LED light source adjustments are also controlled on t...

Страница 26: ...calendar analog to digital converters and many other circuits for signal processing and control The ambient pressure and chassis temperature sensors are also located on this board The main controller...

Страница 27: ...ilable to connect with External I O Port The analog digital port sends and receives analog digital signals to other devices These signals are commonly used to activate gas calibrators or for warning a...

Страница 28: ...ning alarms such as no flow sample mode etc CAUTION Exceeding 12 VDC or drawing greater than 400 mA on a single output or a total greater than 2 amps across the eight outputs can permanently damage th...

Страница 29: ...ith the following steps 1 Undo the screws located in the rear panel 2 Open the chassis lid by releasing the latch pressing the button located on the front panel in the top left hand corner slide the l...

Страница 30: ...esults the instrument should be located in a temperature and humidity controlled environment air conditioned shelter An enclosure temperature of 25 27 C is optimum Whether in a rack or placed on a ben...

Страница 31: ...two meters in length with inch ID and inch OD Sample inlet pressure should not exceed 5 kPa above ambient pressure Tubing must be cut squarely and any burrs removed Remove the inlet port nut and inser...

Страница 32: ...The exhaust must be a suitable distance from the sample inlet to avoid influencing the ambient measurements 2 3 2 Power Connections CAUTION When connecting the mains power to the instrument the follow...

Страница 33: ...p 1 Open the lid and ensure the USB memory stick is installed refer to Figure 6 2 Check the battery is turned on at the main controller PCA refer to Figure 7 3 Turn on the instrument and allow the war...

Страница 34: ...plugs into the pneumatic connections 6 Remove the USB memory stick and pack with instrument refer to Figure 6 7 Place the instrument back into a plastic bag with desiccant packs and seal the bag idea...

Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...

Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...

Страница 37: ...arily operates in the sample measure cycle continuously A background is preformed after warm up and then normally once every 24 hours by default or as defined in the Measurement Settings Menu which ca...

Страница 38: ...Dates Times Numbers etc On the home screen these buttons are used for adjusting the screen contrast Press and hold the up button to increase contrast press and hold the down button to decrease Keypad...

Страница 39: ...oning An orange light indicates there is a minor problem with the instrument but the instrument may still take measurements reliably A green light indicates that the instrument is working and there ar...

Страница 40: ...menu buttons at the bottom of the screen Concentration Units 6 The instrument units are displayed in the bottom right hand corner of the display USB Detection 7 A USB symbol will be displayed in the...

Страница 41: ...wish to view events Instrument This field allows the instrument to be set to either Online normal instrument operation or In Maintenance data is flagged as invalid Safely Remove USB Stick Always selec...

Страница 42: ...r Power Failure This field displays the time and date of the last power failure or when power was disconnected from the instrument 3 4 4 Status Menu Main Menu Analyser State Menu Status Menu The Statu...

Страница 43: ...on 3 4 16 Pressure Calibration Error if the user is performing a pressure calibration Flow Calibration Internal Pump Option Error if the user is performing a flow calibration Flow Fault Ok when the in...

Страница 44: ...Option Current pressure within the flow block Flow Units Select the units that the sample flow will be displayed in slpm or L min Sample Flow Indicates the gas flow through the sample port of the ins...

Страница 45: ...Avg which displays both instantaneous and average concentration on the Home Screen The average is measured over the time period set in Measurement Settings Menu refer to Section 3 4 9 Char 0 has Slash...

Страница 46: ...med Mode refer to Section 3 4 10 2 Select the Cal Type field and select either Timed or Manual Timed mode is an automatic calibration controlled by the Interval between cycles Length of each calibrati...

Страница 47: ...so that a zero calibration can be performed Data is flagged as zero data Span This mode will take air through the calibration port so that a span calibration can be performed Data is flagged as span d...

Страница 48: ...is item will open a dialog box of instructions Pressure Units Select the units that the pressure will be displayed in torr PSI mBar ATM or kPa Ambient The current ambient pressure The current ambient...

Страница 49: ...fer to Section 3 4 17 Pres Temp Flow Comp On default Is used to compensate instrument measurements for environmental fluctuations that might affect readings pressure temperature and flow Off Is used o...

Страница 50: ...low Control field is set to MANUAL Coarse Internal pump speed control Coarse This field is only editable when the Flow Control field is set to MANUAL Fine Internal Pump speed control Fine This field i...

Страница 51: ...low from NC to C drawing sample from the Calibration Port Pressurised Zero Optional Indicates if the optional pressurised zero port valve is Off or On refer to Section 8 4 Pressurised Span Optional In...

Страница 52: ...ent so for a 4 1 dilution ratio enter a value of 4 Enter the ratio here so the instrument can display the correct value on the instrument The default value is 1 00 this indicates no dilution is applie...

Страница 53: ...an log a maximum of 12 parameters Data Log Interval Displays the interval at which the data is saved to the USB memory stick Selecting a 1 sec interval may result in occasional measurements not being...

Страница 54: ...otely as parameters 3 to 5 CAUTION Exceeding these voltages can permanently damage the instrument and void the warranty Input 1 2 3 The sections below are repeated for each analog input Multiplier The...

Страница 55: ...nu These items appear when Output Mode is set to Voltage Voltage Offset Choices are 0V 0 25V or 0 5V This sets the voltage for a reading of 0 Since the output cannot go negative this offset can be use...

Страница 56: ...Low Active High means that the event will be triggered when the line is pulled to 5V Active Low means that the event will be triggered when the line is pulled to 0V Example Here is a typical configura...

Страница 57: ...er Active Pwr Supply Fail Power supply fault A2D Fail Analog to digital fault Flow Fail Sample flow fault Cell Heat Fail Cell heater fault Flow Heat Fail Flow block heater fault Chassis Tmp Fail Chass...

Страница 58: ...strument an IP address IP Address Read or Set mode This is the current IP address of the instrument Netmask Read or Set mode This is the subnet mask of the network the instrument is connected to Gatew...

Страница 59: ...and supervision of qualified service personnel Language Select a language Hardware Menu Refer to Section 3 4 31 Service Displays When set to On new items appear on many different menus These fields ar...

Страница 60: ...istently Default is Aluminium Network Port When Enabled indicates the instrument has a network port installed Default is Disabled Orifice Size Specify the input orifice Default is 0 62 Internal Pump W...

Страница 61: ...d can be connected to DTE Data Terminal Equipment such as a data logger or computer Port 2 also supports a multidrop arrangement a configuration of multiple instruments connected via the same RS232 ca...

Страница 62: ...ction to a laptop running Ecotech s Airodis software to download logged data and remotely control the instrument Due to the nature of USB this is a less reliable permanent connection as external elect...

Страница 63: ...rk Communication optional Instruments with the optional network port installed can be accessed using a TCP IP connection Figure 12 shows examples of some possible configurations for remote access Figu...

Страница 64: ...dit Netmask Change the Netmask to the setup specified by the Modem Router 6 Edit Gateway Change the Gateway to the setup specified by the Modem Router Figure 13 Example of Network Menu Setup 7 Once co...

Страница 65: ...Forwarding Example 4 3 3 Setup Airodis to Communicate with Serinus LAN Below is an example of Airodis setup for a LAN network Ensure the IP address is set to the same as on the instrument Network Menu...

Страница 66: ...ctivate gas calibrators or for warning alarms 4 4 1 Analog Outputs The instrument is equipped with one to three analog outputs that can be set to provide either voltage 0 5V 0 25 5V 0 5 5V 0 10V or cu...

Страница 67: ...to the ground pin 24 and the relevant output pin pin 10 4 Edit 0 5V Calibration until the multimeter reads 0 500 0 002 Accept 5 Edit 5 0V Calibration until the multimeter reads 5 00 0 002 Accept 4 4 1...

Страница 68: ...jumper JP1 on the back panel PCA refer to Figure 17 4 4 4 Digital Status Outputs The instrument is equipped with eight open collector outputs which will convey instrument status condition warning ala...

Страница 69: ...ers already setup in the internal data logger These select few parameters have been chosen for their relevance in assisting in troubleshooting the instrument 4 5 1 Configure Instrument Internal Loggin...

Страница 70: ...tions do not require additional drivers If you wish to connect using a USB cable the driver must first be installed 4 6 1 1 Connecting Over USB If you wish to connect using USB you will need to first...

Страница 71: ...rinus USB driver is located on the green Ecotech resources USB stick under Drivers Ecotech Analyser Select this directory and click Next Figure 21 Update Driver Popup Directory Location If you receive...

Страница 72: ...1 2 Connecting Over Serial RS 232 The following steps outline how to setup the instrument for connection to a PC or datalogger refer to Section 3 4 23 Procedure 1 Open Main Menu Communication Menu Ser...

Страница 73: ...from the instrument using either a full retail paid version of Airodis or with the demo version which is included on the green Ecotech resources USB stick The demo version has limited functionality b...

Страница 74: ...on for each The client may prompt to register with Ecotech or install an update Follow the prompts if it does 3 Once the Client application has loaded click Home Add Station New Physical Station Figur...

Страница 75: ...rument will need to be adjusted manually for DST it will not happen automatically Database Name This is the name to be used for the table in the SQL database containing this station s data It must be...

Страница 76: ...If there was an error connecting to the instrument a red dot will appear next to the station name in the station list on the far left hand side Hovering over the red dot will present you with an erro...

Страница 77: ...e window You can also monitor the status of the download from the Home tab Figure 30 Download Data Status 9 Data will become available in the data manager as it is downloaded You can load data for a d...

Страница 78: ...ll allow you to save your data in CSV format which can be loaded into another program such as Microsoft Excel It is also possible to copy paste Ctrl C Ctrl V data directly from the Airodis data manage...

Страница 79: ...lows the user to Completely control the instrument using a remote screen displayed on the device Download logged data and take snapshots of all the instrument parameters Construct graphs from logged d...

Страница 80: ...format of this menu may vary 4 7 2 Connecting to the Instrument Procedure 1 Open Main Menu Communications Menu Bluetooth Menu to find the Bluetooth ID and PIN refer to Section 3 4 29 2 Touch the Scan...

Страница 81: ...umber pad all button functions actions can be performed by touching the screen This includes the selection buttons and the scroll buttons Touching any part of the screen where there is not already a b...

Страница 82: ...ow Hide NumPad Show or hide the number pad Real Time Plot Refer to Section 4 7 4 Download Refer to Section 4 7 5 Get Parameters Refer to Section 4 7 6 Preferences Refer to Section 4 7 7 4 7 4 Real Tim...

Страница 83: ...collection to set the interval Clear Clears the window and restarts the graphing Save Saves an image of the graph and accompanying data in the location specified in preferences refer to Section 4 7 7...

Страница 84: ...asks to send the saved comma separated text file csv as an attachment to an email This file format does not include the parameter headings just the values Send E Mail Sends an email with the parameter...

Страница 85: ...il formatted as displayed Preferences Refer to Section 4 7 7 4 7 7 Preferences The Preferences Menu allows the operator to adjust the directory settings logged data format and the colour scheme settin...

Страница 86: ...Serinus 60 User Manual 3 0 Page 86 Colour Theme Settings Allows the user to choose a colour scheme for the remote screen either Matrix Classic Emacs or Custom Figure 42 Colour Theme Settings...

Страница 87: ...ms 5 1 Overview Figure 43 Example of a Calibration System CAUTION All calibration gases must be supplied at ambient pressure to avoid damaging the instrument If direct gas cylinder connection is requi...

Страница 88: ...the instrument response across its operating range In general terms the calibration process includes the following steps 1 Establish a reliable and stable calibrating source 2 Provide a satisfactory...

Страница 89: ...calibration first 5 2 1 Full Pressure Calibration This section outlines how to perform a full pressure calibration Using the required equipment follow the steps below to complete a full pressure calib...

Страница 90: ...sure adjustment is required Note Ensure that the instrument has been running for at least one hour before any calibration is performed to ensure the instrument s stability Note Ensure units of measure...

Страница 91: ...uence on hold normal sampling will be interrupted 3 Edit Vacuum Set Pt Read displayed instructions OK 4 Connect a barometer to the Sample Port refer to Figure 5 5 Wait 2 5 minutes and ensure the press...

Страница 92: ...ctions outlined in the next three sections Note Ensure the instrument has been running for at least one hour before any calibration is performed to ensure sufficient stability 5 4 1 Calibration Port E...

Страница 93: ...rformed at the upper end of the instrument s measurement range Ecotech recommends calibration at 80 of the full scale measurement or operating range of the instrument While the instrument range is com...

Страница 94: ...ual Accept 5 Select Cal Mode Measure Accept 6 Let the instrument stabilise typically 15 minutes 7 Enter Span Calibrate NO2 Enter the span output concentration Accept 8 Disconnect the span source and r...

Страница 95: ...gs against your local applicable standards If an instrument fails a span precision check based on the users local applicable standards perform a span calibration refer to Section 5 5 If an instrument...

Страница 96: ...ter plot of expected concentration versus the recorded instrument response and use linear regression to calculate the line of best fit and the correlation factor R2 refer to the users local applicable...

Страница 97: ...ting 5 8 High Pressure Zero Span Valve If the instrument was ordered with this option the internal pressurised calibration valves will already be installed within the instrument as either a zero or sp...

Страница 98: ...eter Gas Cylinder Procedure 1 Ensure the gas cylinder is fitted with an appropriate gas regulator with a shut off valve 2 Connect a line of stainless steel tubing between the gas cylinder and the Aux...

Страница 99: ...bration Port is between 0 5 and 1 slpm This flow rate is you excess calibration gas Note Do not exceed a pressure of 2 bar this can damage the instrument and cause gas leakage Return to Normal Operati...

Страница 100: ...equired Calibrated Flow Meter Gas Cylinder of Zero air Gas Cylinder of NO2 Procedure 1 Ensure the gas cylinder is fitted with an appropriate gas regulator with a shut off valve 2 Connect a line of sta...

Страница 101: ...bration gas Note Do not exceed a pressure of two bars this can damage the instrument and cause gas leakage 10 Select Cal Mode Span Accept 11 Adjust the regulator pressure until the flow meter on the v...

Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...

Страница 103: ...1078 E031079 Flow Meter Select Range Range 50 sccm to 5000 sccm PN ZBI 200 220M Range 300 sccm to 30000 sccm PN ZBI 200 220H Minifit Extraction Tool PN T030001 Orifice Sintered Filter Extraction Tool...

Страница 104: ...liance standards that differ in various regions The following is recommended by Ecotech as a guide Compliance with local regulatory or international standards is the responsibility of the user Table 6...

Страница 105: ...nsure it is less than 600 Mm 1 Observe the trend and if the reading will increase to 650 Mm 1 within the next 6 months perform a background scrubber DAU check 6 3 1 Perform NO2 span calibration 5 5 Pe...

Страница 106: ...l Mode Zero Accept 5 Allow the instrument to sample Zero air for approximately five minutes 6 Record the instrument response as the Initial value Note This value should be 0 00 1 ppb If it is not a ba...

Страница 107: ...ounter clockwise 4 Remove the filter plunger from the casing place finger on tubing connector and pull to the side refer to Figure 52 Figure 52 Removing the Plunger 5 Remove the old filter paper wipe...

Страница 108: ...ernal pump if an internal pump is installed refer to Section 8 2 3 and allow the instrument to return to ambient 2 Loosen the Kynar nuts from both ends of the sample DFU 3 Remove and replace the sampl...

Страница 109: ...eak Test Jig PN H050069 Kynar Blocker Nuts Tubing and Assorted Fittings 5 8 Spanner 9 16 Spanner Procedure Note Ensure that the instrument has been running for at least one hour before this procedure...

Страница 110: ...o return to ambient Inspect the internal tubing to ensuring that the tubing is cleanly connected to the fittings and that the internal Teflon lining has not been kinked or crumpled 13 Remove the leak...

Страница 111: ...al Pump Menu 20 Select Flow Control Start Accept 6 3 7 Clean Pneumatics The calibration valve manifold will require disassembling and cleaning Ideally the valves and the manifold should be cleaned in...

Страница 112: ...ment may have a hardware fault The cell pressure PCA and the manifold pressure PCA have test points To determine if the pressure sensor is faulty simply measure the voltage on the test points shown in...

Страница 113: ...l the following screen appears Ecotech Serinus Analyser V3 1 Bootloader Press 1 to enter Bootloader If the instrument displays the normal start up screen the power will need to be toggled and another...

Страница 114: ...that firmware is placed in a folder called FIRMWARE in the front panel USB Port 3 Enter the Bootloader refer to Section 6 4 4 Select option 3 upgrade from USB memory stick press 3 on the keypad 5 Wai...

Страница 115: ...ent is available for excess gas Leaks A leak in the instrument or calibration system dilutes the sample stream and causes low span readings and noise Refer to Section 6 3 6 Measurement cell heater A f...

Страница 116: ...eck valves Re calibrate pressure sensors Unstable flow or pressure readings Faulty pressure sensors Check pressure transducer calibration Check calibration valve block is functioning and not blocked I...

Страница 117: ...Page 117 Error Message Problem Cause Solution Faulty zero air Ensure zero air source is not overly polluted Leak Perform leak test Refer to Section 6 3 6 7 1 Leak Check Figure 59 Leak Check Troublesho...

Страница 118: ...Serinus 60 User Manual 3 0 Page 118 7 2 Flow Fault Figure 60 Zero Flow Fault Troubleshooting Flow Chart Section 2 3 1 Section 5 2...

Страница 119: ...Troubleshooting Page 119 7 3 Noisy Unstable Readings Figure 61 Noisy or Unstable Readings Troubleshooting Flow Chart Section 6 3 6 Section 5 2...

Страница 120: ...Serinus 60 User Manual 3 0 Page 120 7 4 Flow Block Temperature Failure Figure 62 Flow Block Temp failure Troubleshooting Flow Chart...

Страница 121: ...Troubleshooting Page 121 7 5 Measurement Cell Temperature Failure Figure 63 Measurement Cell Temperature Failure Troubleshooting Flow Chart...

Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...

Страница 123: ...r State Menu 2 Variant take note 3 Ecotech ID take note 4 Board Revision take note 5 Firmware Ver take note 6 Open Main Menu Service Menu 7 Save Save Configuration CONFIG CFG Accept 8 Save Save Parame...

Страница 124: ...and access the files 11 Best practice is to email all the on the USB memory stick but if it s to large just send 12 The CONFIG CFG and PARAM TXT files that are saved in the CONFIG folder 13 The LOG f...

Страница 125: ...n the pneumatic diagram dashed line and requires no operational changes to the instrument Figure 66 Dual Filter Option Installed 8 2 Internal Pump PN E020124 The Serinus internal pump installed option...

Страница 126: ...installing the Serinus into a 19 rack the Serinus is 4RU in height Table 8 Included Parts Rack Mount Kit Parts Removed Quantity Part Numbers Rack Slide Set 1 H010112 Rack Mount Adaptors 4 H010133 Rac...

Страница 127: ...nner slide rails to each side of the instrument using M4 x 10 button screws three on each side refer to Figure 68 Figure 68 Assemble Inner Slide on Chassis 4 Install rack mount ears on the front of th...

Страница 128: ...outer slide rails using M4 x 10 button screws washers and locknuts Do not fully tighten at this stage as minor adjustments will be required to suit the length of the rack refer to Figure 70 Figure 70...

Страница 129: ...des to Front of Rack 8 Now carefully insert the instrument into the rack by fitting the instrument slides into the mounted rails Ensuring that the rack slide locks engage on each side you will hear a...

Страница 130: ...lock labelled Push and push it in whilst sliding the instrument out of the rack complete this for both sides while carefully removing instrument Figure 73 Slide Clips 8 4 High Pressure Zero Span Valv...

Страница 131: ...n Tubing 1 4 OD 1 8 ID Clear 3 Feet T010011 Neoprene Tubing 1 4 OD 1 8 ID Black 3 Feet T010021 9 2 Maintenance Kit This maintenance kit is required when performing annual maintenance on the instrument...

Страница 132: ...labour from date of shipment the warranty period The warranty period commences when the product is shipped from the factory Lamps fuses batteries and consumable items are not covered by this warranty...

Страница 133: ...0047 Interface PCA C010048 Display Ribbon Cable C020065 Rear Panel PCA Ribbon Cable C020066 Heater Extension Cable C020094 Bluetooth Cable C020119 Interface Cable C020136 Mixer Preamp Cable C020137 Ad...

Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...

Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...

Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...

Страница 137: ...Parts List Page 137...

Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...

Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...

Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...

Страница 141: ...Serial ID assigned in the Main Menu Communications Menu Serial Communication Menu ETX ASCII End of Text 0x03 hex Checksum The XOR of the individual bytes except for STX ETX EOT and Checksum Message l...

Страница 142: ...8 137 137 0 137 137 0 137 137 A 2 Commands A 2 1 Communication Error Where Command byte 0 Message byte 1 0 Message byte 2 0 7 If the command byte of a response is 0 this indicates an error has occurre...

Страница 143: ...essage field of 50 51 52 to a Serinus S40 would return a 15 byte message as shown below Table 18 Example Get IEEE Response data 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 50 NO reading 51 NOx reading 52 NO2...

Страница 144: ...o an individual instrument Table 19 Advanced Protocol Parameter List Description Notes 1 Cal Zero Valve 0 Zero 1 Cal 2 Internal Span Valve 0 Off 1 On 3 Positive Supply Positive analog supply voltage 4...

Страница 145: ...block pressure 30 Analog input 5 Cell pressure 31 Analog input 6 Ambient pressure 32 Analog input 7 Raw ADC calibration input 33 Analog input 8 Reserved 34 Analog input 9 Concentration data 35 Analog...

Страница 146: ...55 Gas 3 Avg Average of the readings for Gas3 of the last n minutes where n is the averaging period 56 Instrument Gain Current calibration value default is 1 0 57 Serial ID Multidrop or Bayern Hessen...

Страница 147: ...sets of values Set Calibration State 0 MEASURE 1 CYCLE 2 ZERO 3 SPAN Get IEEE Value 0 MEASURE 1 ZERO 2 SPAN 86 Primary Raw Conc For S40 before NOx background and gain 87 Secondary Raw Conc Only for mu...

Страница 148: ...lt AO 1 5 0 Volt Calibration value 102 F Scale Volt AO 2 103 F Scale Volt AO 3 104 Z Adj Volt AO 1 0 5 Volt Calibration value 105 Z Adj Volt AO 2 106 Z Adj Volt AO 3 107 Negative Supply Negative analo...

Страница 149: ...BACKGROUND FILL ZERO 18 BACKGROUND MEASURE ZERO 19 SPAN FILL 20 SPAN MEASURE 21 SPAN FILL AUX 22 SPAN MEASURE AUX 23 SPAN FILL AUX2 24 SPAN MEASURE AUX2 25 BACKGROUND FILL SPAN 26 BACKGROUND MEASURE S...

Страница 150: ...cient D 128 Backgnd Pause Time In seconds 129 Bkgnd Intrleav Fact 130 Cal Pressure 2 Calibration Pressure for 2nd gas 131 2nd Instrument Gain Unused always reports 1 0 132 Background voltage Units in...

Страница 151: ...tion Factor for first gas 152 PTF Correction gas 2 Pressure Temperature Flow Compensation Factor for second gas in dual gas analysers 153 Inst Cell Pressure Instantaneous cell pressure 154 Manifold Ma...

Страница 152: ...IVE FILTER 165 NO2 Filter enabled 0 Disabled 1 Enabled 166 Background Interval 0 24 Hrs 1 12 Hrs 2 8 Hrs 3 6 Hrs 4 4 Hrs 5 2 Hrs 6 Disable 167 Service COM1 Baud Serial baud rate 0 1200 bps 1 2400 bps...

Страница 153: ...Cell Press gas5 Cell Pressure for Gas 5 Nx 183 Gas 4 Inst Calculated gas concentration currently displayed on front screen E g NH3 184 Gas 4 Avg Average of the readings for Gas 4 of the last n minutes...

Страница 154: ...0 EC9800 1 Bayern Hessen 2 Advanced 3 Modbus 208 Gas 4 Offset A offset applied to Gas 4 209 O3 GEN Fine Pot Ozone generator control DAC controlled DAC 0 64535 210 O3 Gen Lamp Current Units in mA 211...

Страница 155: ...I D can be replaced with the string B 2 Commands B 2 1 DCONC Function Sends the current instantaneous concentration data to the serial port Format DCONC DEVICE I D TERMINATOR Device response GAS SPACE...

Страница 156: ...ICE I D TERMINATOR Device response ACK if the instrument is able to perform the command NAK if not B 2 4 ABORT Function Commands the instrument to abort the current span zero mode and return to measur...

Страница 157: ...ation of block check value LSB The block check algorithm begins with 0 and exclusive OR s each ASCII character from STX to ETX inclusive This block check value is converted to ASCII format and sent af...

Страница 158: ...Check Operation Character Hex Value Binary Block Check STX 02 0000 0010 0000 0010 S 53 0101 0011 0101 0001 T 54 0101 0100 0000 0101 8 38 0011 1000 0011 1101 4 34 0011 0100 0000 1001 3 33 0011 0011 001...

Страница 159: ...e number of gasses reported 0 5 The text in between the will be repeated once for each gas reported kkk The Bayern Hessen instrument ID nnnn ee Gas concentration ss Status byte see table below for ind...

Страница 160: ...5 Cell heater failure S30 S40 and S50 family only 6 7 C 2 2 ST Set the instrument mode Command Format STX ST kkk SP command ETC bcc1 bcc2 Where kkk Device s Serial ID This field is optional but if pr...

Страница 161: ...mand may request from 2 to 124 registers which is to say from 1 to 62 values The first index is specified by Start register all following indexes are in sequential order To read values that are not se...

Страница 162: ...of the four bytes Little Endian means it is at the left CRC As general command format D 2 2 Write Holding Register The only supported use for this command is to set the instrument into a calibration...

Страница 163: ...he initiating command s function code 128 so either 131 read or 144 write Exception code The error code see table below CRC As general command format Table 24 Modbus Error Codes Value Error 1 Illegal...

Страница 164: ...change without notice Images used are for illustrative purposes only All trademarks and registered trademarks are the property of their respective owners 2022 Acoem and all related entities All right...

Отзывы: