![ACOEM Met One Instruments GAS-1060 Скачать руководство пользователя страница 141](http://html1.mh-extra.com/html/acoem/met-one-instruments-gas-1060/met-one-instruments-gas-1060_user-manual_459535141.webp)
Parts List
Page 141
Appendix A.
Advanced Protocol
The Advanced protocol allows access to the full list of instrument parameters.
A.1
Command Format
All commands and responses sent to and from the Instrument will be in the following packet format
to ensure data is reliable.
Table 13 – Packet Format
1
2
3
4
5
6 … 5+n
6+n
7+n
STX (2)
Serial ID
Command
ETX (3)
Message
Length (n)
Message
Checksum EOT (4)
Where:
<STX>
ASCII Start of Text = 0x02 hex.
Serial ID
The Serial ID assigned in the Main Menu
Communications Menu
Serial Communication Menu.
<ETX>
ASCII End of Text = 0x03 hex.
Checksum
The XOR of the individual bytes except for STX, ETX, EOT and
Checksum.
Message length
Must be in the range 0 to 32. Responses from the instrument can have
a message Length of 0 to 255.
<EOT>
ASCII End of Transmission = 0x04 hex.
Examples
A basic request for Primary gas data would be as follows:
Table 14 – Example: Primary Gas Request
Byte Number
1
2 3
4
5
6
7
8
Description
STX
ID Command ETX
Message
Length
Primary Gas
Conc
Checksum
EOT
Value
2
0 1
3
1
50
50
4
Checksum
Calculation
0 0
1=1
1
1=0
0
50=50
50
And a sample response:
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...