Performing a zero calibration will adjust the
Zero Offset NO2
. This offsets can be checked in the
Main Menu
Service Menu
Calculation Factors Menu
and should be very close to zero. A large
offset may indicate a problem with the instrument or calibration system (refer to Section 7).
Note:
Ecotech encourages regular zero checks; however Ecotech recommends that the zero
calibration only be performed when specifically required as it may mask issues that should be
addressed during maintenance/service.
A zero calibration can be performed either through the
Calibration Port,
Sample Port
or
Internal
zero
. Refer to the instructions outlined in the next three sections:
Note:
Ensure the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
5.4.1
Calibration Port
Equipment Required
Zero source
Procedure
1.
Ensure a suitable zero source is connected to the
Calibration port
.
2.
Open -
Main Menu
Calibration Menu.
3.
Select -
Cal. Type
Manual
- Accept.
4.
Select -
Zero Source
External
- Accept.
5.
Select -
Cal. Mode
Zero
- Accept.
6.
Allow the instrument time to achieve a stable response.
7.
Enter -
Zero Calibrate NO2
- OK.
8.
Select -
Cal. Mode
Measure
- Accept (To return to sample measure).
5.4.2
Sample Port
Equipment Required
Zero source
Procedure
1.
Ensure a suitable zero source is connected to the
Sample Port
.
2.
Open -
Main Menu
Calibration Menu.
3.
Select -
Cal. Type
Manual
- Accept.
4.
Select -
Zero Source
External
- Accept.
5.
Select -
Cal. Mode
Measure
- Accept.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...