Figure 58 – Typical Test Point Reading of Measurement Cell Pressure Sensor
6.4
Bootloader
The Serinus Bootloader is the initial set of operations that the instruments’ microprocessor performs
when first powered up (similar to the BIOS found in a personal computer). This occurs every time the
instrument is powered up or during instrument resets. Once the instrument boots up, it will
automatically load the instruments’ firmware. A service technician may need to enter the Bootloader
to perform advanced microprocessor functions as described below.
To enter the Bootloader turn off the power to the instrument. Press and hold the plus key while
turning the power on. Hold the Plus key until the following screen appears.
** Ecotech Serinus Analyser **
V3.1 Bootloader
Press ‘1’ to enter Bootloader
If the instrument displays the normal start up screen, the power will need to be toggled and another
attempt will need to be made to enter the Bootloader. Once successful, press 1 to enter the
Bootloader Menu
.
6.4.1
Display Help Screen
Once in the Bootloader screen it is possible to redisplay the help screen by pressing 1 on the keypad.
6.4.2
Communications Port Test
This test is very useful for fault finding communication issues. It allows a communication test to be
carried out independent to any user settings or firmware revisions.
This command forces the following communication ports to output a string of characters: Serial Port
RS232 #1, USB rear and Ethernet Port. The default baud rate is 38400 for the RS232 Serial Port. To
initiate the test press the number 2 key from the Bootloader screen.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...