Serinus 60 User Manual 3.0
Page 42
Communications Menu
Refer to Section 3.4.21.
3.4.3
Analyser State Menu
Main Menu
Analyser State Menu
This displays the status of various parameters that affect instrument measurements.
Status Menu
Refer to Section 3.4.4.
Temperature Menu
Refer to Section 3.4.5.
Pressure & Flow Menu
Refer to Section 3.4.6.
Voltage Menu
Refer to Section 3.4.7.
Model
This field will always display Serinus.
Variant
The variant of the Serinus model (e.g. S60).
Range
The range of the Serinus model (Standard, High or Trace).
Ecotech ID
The Ecotech ID number.
Serial No.
The main controller PCA serial number.
Board Revision
The main controller PCA version.
Firmware Ver.
This field displays the firmware version currently in use on this
instrument. This can be important when performing diagnostics
and reporting back to the manufacturer.
Power Failure
This field displays the time and date of the last power failure or
when power was disconnected from the instrument.
3.4.4
Status Menu
Main Menu
Analyser State Menu
Status Menu
The
Status Menu
presents a list of the current
Pass/Fail
statuses of the main components.
During warm-up, the status of some parameters will be a dashed line.
Event Log
This field enters a screen with a log of all the events that the
instrument has performed. These events include errors and
warnings. This log is stored on the USB memory stick.
The log is organised by month. When you enter this screen you will
be prompted to enter the month for which you wish to view
events.
Show Error List
This field allows the user to display the list of current errors and
warnings on the screen.
Next Service Due
This field is visible with the next service due date if the service is
due within the next two weeks.
+5V Supply
Pass if the +5 V power supply is within the acceptable range.
+12V Supply
Pass if the +12 V power supply is within the acceptable range.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...