Calibration
Page 89
Regulatory agencies establish the time intervals at which the instrument must be calibrated to
ensure satisfactory data for their purposes.
5.2
Pressure Calibration
The pressure sensors are a vital component of the instrument operation and the pressure calibration
should be checked on installation or whenever maintenance is performed.
A thorough leak check must be performed prior to performing a pressure calibration (refer to Section
6.3.6).
The pressure calibration can either be a two point calibration (one point under vacuum and another
point at ambient pressure) or a single ambient point calibration when very minor adjustments are
required.
Note:
Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure the instrument’s stability. When performing a two point pressure calibration, it
is advisable to perform the vacuum pressure calibration first.
5.2.1
Full Pressure Calibration
This section outlines how to perform a full pressure calibration. Using the required equipment follow
the steps below to complete a full pressure calibration.
Note:
This procedure is for firmware version 3.44.0001 and above. Ensure that the instrument has
been running for at least one hour before the calibration is performed.
Note:
Ensure units of measure are the same on both the barometer and instrument.
Equipment Required
Barometer
Vacuum source
Procedure
1.
Disconnect all external tubing connected to the rear ports of the instrument.
2.
Open -
Main Menu
Calibration Menu
Pressure Calibration Menu
- (read note) - OK.
Note:
This action will place the valve sequencing on hold; normal sampling will be interrupted.
3.
Edit -
Vacuum Set Pt.
- (Read displayed instructions) - OK.
4.
Connect a barometer to the
Sample Port
(refer to Figure 5).
5.
Connect a vacuum source to
Exhaust Port
(refer to Figure 5).
6.
Wait 2-5 minutes and ensure the pressure reading on the barometer has dropped and is stable.
7.
Enter the barometer reading into the instrument - Accept.
8.
Read displayed instructions - OK.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...