![ACOEM Met One Instruments GAS-1060 Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/acoem/met-one-instruments-gas-1060/met-one-instruments-gas-1060_user-manual_459535109.webp)
Service
Page 109
Procedure
1.
Turn off the external pump (if an internal pump is installed refer to Section 8.2.3) and allow the
instrument to return to ambient.
2.
Loosen the Kynar nuts from both ends of the background DAU.
3.
Remove and replace the background DAU (refer to Figure 54) and tighten the Kynar nuts finger
tight.
Figure 55 – Background DAU
6.3.6
Leak Check
Equipment Required
Source of Vacuum (pump)
Leak Test Jig (PN: H050069)
Kynar ¼” Blocker Nuts
Tubing and Assorted Fittings
5/8” Spanner
9/16” Spanner
Procedure
Note:
Ensure that the instrument has been running for at least one hour before this procedure is
performed.
1.
Disconnect all external tubing connected to the rear ports of the instrument.
2.
Connect a leak check jig to the Exhaust Port of the instrument.
3.
Connect a vacuum source to the shut off valve end of the leak test jig ensuring the shut off valve
is in the open position.
4.
Open -
Main Menu
Service Menu
Diagnostics Menu
Valve Menu
.
5.
Disable -
Valve Sequencing
Disabled
.
6.
Turn
Off
all the valves
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...