![ACOEM Met One Instruments GAS-1060 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/acoem/met-one-instruments-gas-1060/met-one-instruments-gas-1060_user-manual_459535041.webp)
Operation
Page 41
For example, the manual calibration type and mode can only be changed when the instrument has
finished the warm-up process.
3.4.1
Quick Menu
The
Quick Menu
contains all the maintenance tools in one easy to use screen. It allows operators to
perform calibrations, check important parameters and review the service history.
Span Calibrate NO2
This field is used to perform a span calibration and should be
only used when a known concentration of span gas is running
through the measurement cell and the reading is stable.
Activating the span calibrate field for a named gas will open a
dialog box. Enter the concentration of the span gas that the
instrument is sampling and press
Accept
.
Event Log
This field enters a screen with a log of all the events that the
instrument has performed. These events include errors and
warnings. This log is stored on the removable USB memory stick.
The log is organised by month. When you enter this screen you
will be prompted to enter the month for which you wish to view
events.
Instrument
This field allows the instrument to be set to either
Online
(normal instrument operation) or
In Maintenance
(data is
flagged as invalid).
Safely Remove USB Stick
Always select this menu item before removing the USB memory
stick or select the same menu item from the
Service Menu
(refer
to Section 3.4.12). Failure to do this may cause corruption of the
memory stick.
Instrument Gain
This is a multiplication factor which is used to adjust the
concentration measurement to the appropriate level (set by
performing a
Span Calibrate
). This should be recorded after each
calibration in the station log book.
Next Service Due
A field that notifies the user when the next instrument service is
due. This value is editable in the Next Service Due field of the
Advanced Menu (refer to Section 3.4.30). This field is only
displayed in the two weeks prior to the date displayed in this
field or after the date has occurred.
3.4.2
Main Menu
There are six menus on the
Main Menu
screen.
Analyser State Menu
Refer to Section 3.4.3.
General Settings Menu
Refer to Section 3.4.8.
Measurement Settings Menu
Refer to Section 3.4.9.
Calibration Menu
Refer to Section 3.4.10.
Service Menu
Refer to Section 3.4.12.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...