Serinus 60 User Manual 3.0
Page 48
Span Compensation
When
Enabled
the instrument will automatically perform a
Span
Calibrate NO2
at the end of the
Cycle Time
and adjust the gain
based on the
Span Level
.
When
Disabled
it will do a precision check only, no adjustment is
made.
Timed
mode with span compensation enabled does not fulfil U.S.
EPA approval or EN certification.
Span Level
Enter the concentration of span gas expected. Used when the
Span Compensation
is
Enabled.
3.4.11
Pressure Calibration Menu
Main Menu
Calibration Menu
Pressure Calibration Menu
Entering this menu will set the valves to the pressure calibration configuration; leaving the menu will
restore the valves to normal operation (refer to Section 5.2).
Vacuum Set Pt.
The zero point for the calibration. Activating this item will open a
dialog box of instructions.
Ambient Set Pt.
The high point for the calibration. Activating this item will open a
dialog box of instructions.
Pressure Units
Select the units that the pressure will be displayed in (
torr
,
PSI
,
mBar
,
ATM
or
kPa
).
Ambient
The current ambient pressure.
The current ambient pressure displayed as a raw voltage.
Cell
The current pressure in the measurement cell.
The current measurement cell pressure displayed as a raw voltage.
3.4.12
Service Menu
Main Menu
Service Menu
Diagnostics Menu
Refer to Section 3.4.13.
Calculation Factors Menu
Refer to Section 3.4.20.
Load Auto-Backup Config.
Loads the auto-backup configuration file. The configuration is
automatically backed up every night at midnight.
Load Configuration
Loads a user selectable configuration file from the USB memory
stick.
Save Configuration
Saves all of the EEPROM-stored user-selectable instrument
configurations to the USB memory stick (calibration and
communication settings, units, instrument gain, etc.). If there are
problems with the instrument use this function to save settings to
the USB memory stick and send this file (together with the
parameter list save) to your supplier with your service enquiry.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...