Operation
Page 49
Save Parameter List
Saves a text file of various parameters and calculation factors. If
you have problems with the instrument use this function to save
settings to the USB memory stick and send this file (together with
the configuration save) to your supplier with your service enquiry.
Instrument
This field allows the instrument to be set to either
Online
(normal
instrument operation) or
In Maintenance
(data is flagged as
invalid).
Next Service Due
Displays when the next scheduled service is due.
Safely Remove USB Stick
This command must be activated to safely remove the USB
memory stick.
System Restart
Activating this will restart the instrument.
3.4.13
Diagnostics Menu
Main Menu
Service Menu
Diagnostics Menu
Digital Pots Menu
Refer to Section 3.4.14.
Internal Pump Menu
[Internal Pump Option]
Refer to Section 3.4.15
Valve Menu
Refer to Section 3.4.16.
Tests Menu
Refer to Section 3.4.17.
Pres/Temp/Flow Comp.
On
(default): Is used to compensate instrument measurements for
environmental fluctuations that might affect readings (pressure,
temperature and flow).
Off
: Is used only when running diagnostics.
Control Loop
Enabled
(default): Allows the instrument to automatically adjust
digital pots and other outputs.
Disabled
: Prevents the instrument from changing most outputs so
the service technician can manually control them.
3.4.14
Digital Pots Menu
Main Menu
Service Menu
Diagnostics Menu
Digital Pots Menu
Digital pots are electronically controlled digital potentiometers used for adjustments to operations of
the instrument. Each of the digital pots can go from 0-255. This menu should be accessed only during
diagnostics.
Unless the
Control Loop
is
Disabled
(refer to Section 3.4.13), changes to the pots may be modified by
the instrument. This is intentional; some diagnostics are best done with instrument feedback and
some are best done without.
Extinction
Mm-1
A measure of light scattering and absorption of NO
2
.
Signal Voltage
0.1 - 2.5V
Signal voltage reading
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...