Serinus 60 User Manual 3.0
Page 50
Meas. Count
0 - 15
This value increases by 2 every 5 seconds.
Sig. Count 1
0 - 16777215
A raw intensity count of the response during the first 90° of the signal
phase.
Sig. Count 2
0 - 16777215
A raw intensity count of the response during the second 90° of the
signal phase.
Sig. Count 3
0 - 16777215
A raw intensity count of the response during the third 90° of the
signal phase.
Sig. Count 4
0 - 16777215
A raw intensity count of the response during the final 90° of the signal
phase.
LED Drive
0-255
Controls the current used to drive the LED thus varying the intensity.
3.4.15
Internal Pump Menu (Option)
Main Menu
Service Menu
Diagnostics Menu
Internal Pump Menu
Flow Control
Set to
MANUAL
to disable the automatic flow control.
AUTO
allows
the flow PID to modify the pump coarse and fine settings.
START
will
transition to
AUTO
after one second.
Internal Pump
This field allows the internal pump to be turned on or off. This field
is only editable when the
Flow Control
field is set to
MANUAL
.
Coarse
Internal pump speed control (Coarse). This field is only editable
when the
Flow Control
field is set to
MANUAL
.
Fine
Internal Pump speed control (Fine). This field is only editable when
the
Flow Control
field is set to
MANUAL
.
Sample Flow
Current gas flow. This is only accurate when reading close to the
flow calibration point.
Internal Pump
The current pressure reading of the flow block which is down stream
of the orifice.
3.4.16
Valve Menu
Main Menu
Service Menu
Diagnostics Menu
Valve Menu
The
Valve Menu
allows the user to observe the instrument controlled switching of the valves. If the
valve is On it means the valve is energised. When a three way valve is in the On state it will now be in
the NC (normally closed) position as shown in the plumbing schematic. When the valve sequencing is
disabled the user has the ability to turn the valve Off and On manually. It is recommended that the
valve menu be used by a trained technician following the plumbing schematic (refer to Section 9.5).
Note:
When interpreting the information below regarding the flow path through the valve note that
(NC = Normally Closed), (NO = Normally Open) and (C = common).
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...