144 Background (hrs)
The background interval in hours.
0.0 if disabled
0.25 for every 15 minutes
0.30 for every 20 minutes
0.50 for every 30 minutes
1.00 for every hour
…
24.00 for once a day
145 Cycle Time
In minutes
146 CO Cooler Pot
CO Cooler voltage adjustment POT
147 CO Source Pot
CO Source voltage adjustment POT
148 CO Test Meas. Pot
Diagnostics use only
149 CO Test Ref. Pot
Diagnostics use only
150 O3 Ref Average
S10 Background Average
151 PTF Correction (gas 1)
Pressure Temperature Flow Compensation Factor
for first gas
152 PTF Correction (gas 2)
Pressure Temperature Flow Compensation Factor
for second gas in dual gas analysers.
153 Inst. Cell Pressure
Instantaneous cell pressure
154 Manifold
Manifold Pressure in S40 instruments
155 Cell Press. (gas1)
Cell Pressure for Gas 1
156 Cell Press. (gas2)
Cell Pressure for Gas 2
157 Cell Press. (Bgnd)
Cell Pressure when in Background
158 Background
0 = the instrument is measuring a gas sample
1 = the instrument is measuring background air
159 Gas To Measure
S51 only; see Measurement Settings Menu
0 = Measure both gasses
1 = Measure SO2 only
2 = Measure H2S only
160 Valve States
Diagnostic use only
161 Temperature Units
0 = "°C"
1 = "°F"
2 = "K"
162 Pressure Units
0 = "torr"
1 = "psi"
2 = "mbar"
3 = "atm"
4 = "kPa"
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...