Serinus 60 User Manual 3.0
Page 46
[Backgrounds]
Time
[Backgrounds]
Edit and display the time that the next background will run. The
time is set using a 24 hour clock.
Repeat
[Backgrounds]
Defines an interval value for the repeat of the background based
on the
Units
selected. This field indicates the delay period; once
the specified amount of time has lapsed the background will
automatically run again. The user can edit this field but some
restriction apply depending on the
Units
selected. Default is “1”.
Units
[Backgrounds]
This is where the user can define the type of units for the
Repeat
delay period. For example: A
Repeat
of “
1
” and
Units
of “
Days”
means that a calibration will automatically run every day at the
defined time. Default is “Days”
3.4.10
Calibration Menu
Main Menu
Calibration Menu
Calibrating the instrument should be done with care (refer to Section 5 before using these menus).
Cal. Type
Depending on the selection in this field, a number of extra menu
items will be displayed. These are separately documented in
Manual Mode
(refer to Section 3.4.10.1) and
Timed Mode
(refer
to Section 3.4.10.2).
Select the
Cal. Type
field and select either
Timed
or
Manual
.
Timed
mode is an automatic calibration controlled by the:
Interval between cycles
Length of each calibration cycle
Time when the calibration will begin
Check only or automatic compensation
Timed calibration with span compensation enabled does not fulfil
U.S. EPA approval.
Manual
mode allows the user to choose the type of calibration
they wish to perform and will open the appropriate valves in
preparation for the user to perform a manual calibration. The
setup used will depend on the
Cal. Mode
selected.
Manual
mode is set as default.
Zero Source
Select whether the instrument will sample from the external
calibration port or from the internal zero source when zero gas is
requested.
Cycle Time
The duration of each
Cal. Mode
(span and zero) when performing
Cycle Mode
(refer to Section 3.4.10.1) or
Cal. Type
is set to
Timed
(refer to Section 3.4.10.2).
Span Calibrate NO2
This field is used to perform a span calibration and should be only
used when a known concentration of span gas is running through
the reaction cell and the reading is stable.
Activating the span calibrate field for a named gas will open a
dialog box. Enter the concentration of the span gas that the
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...