Operation
Page 47
instrument is sampling and press
Accept
.
Zero Calibrate NO2
This command is used to correct the zero calibration setting. This
option should be used only when zero gas is running through the
reaction cell (refer to Section 5 before using this command).
Manual Background
Immediately perform a background.
Pressure Calibration Menu
This menu allows the user to calibrate the pressure sensors (refer
to Section 3.4.11).
Pressure NO2
This field displays the measurement cell pressure measured during
the last calibration.
Temperature
Measurement Cell temperature when the last span calibration was
performed.
3.4.10.1
Manual Mode
These items appear in the
Calibration Menu
when
Cal. Type
is set to
Manual
.
Cal. Mode
When calibration type is set to
Manual
the instruments
operational mode can be chosen from the following:
Measure
: Is the normal measurement through the sample port.
Zero
: This mode will take air through the calibration port so that
a zero calibration can be performed. Data is flagged as zero
data.
Span
: This mode will take air through the calibration port so
that a span calibration can be performed. Data is flagged as
span data.
Cycle
: Performs a zero and a span
Cal. Mode
and then returns
to measure mode. The length of time spent measuring each
calibration mode is set in
Cycle Time
(refer to Section 3.4.10).
While the instrument is still in the warm-up period (refer to
Section 3.1) the
Cal. Mode
cannot be changed from
Measure
mode.
3.4.10.2
Timed Mode
These items appear in the
Calibration Menu
when
Cal. Type
is set to
Timed
.
Date
Enter the date for the next calibration to start.
Time
Enter the time that the calibration will be performed. The time is
set using a 24 hour clock.
Repeat
This field indicates the delay period; once the specified amount of
time has lapsed the calibration will automatically run again. The
user can edit this field (from 1 to 20,000 units).
Units
This is where the user can define the type of units for the
Repeat
delay period. For example: A
repeat
of “
3
” and units of “
Days”
means that a calibration will automatically be performed every
three days.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...