Operation
Page 51
Valve Sequencing
When
Enabled
the valves will turn On and Off under the
instruments control (even if you have manually turned
Off
or
On
a
valve).
When
Disabled
the valves will change only in response to a user’s
action.
Sample/Cal
Indicates if the
Sample/Cal
valve on the
Calibration Valve
Manifold
is
Off
or
On
. This will determine the port the instrument
draws its sample from.
Off
= Flow from NO to C (drawing sample from the
Sample Port
).
On
= Flow from NC to C (drawing sample from the
Sample Port
through the charcoal DAU or from the
Calibration Port
depending
on the state of the
Background /Cal
valve).
Background/Cal
Indicates if the
Background/Cal
valve on the
Calibration Valve
Manifold
is
Off
or
On
. This will determine the port the instrument
draws its sample from.
Off
= Flow from NO to C (drawing sample from the
Sample Port
through the charcoal DAU).
On
= Flow from NC to C (drawing sample from the
Calibration
Port
).
Pressurised Zero
[Optional]
Indicates if the optional pressurised zero port valve is
Off
or
On
(refer to Section 8.4).
Pressurised Span
[Optional]
Indicates if the optional pressurised span port valve is
Off
or
On
(refer to Section 8.4).
3.4.17
Tests Menu
Main Menu
Service Menu
Diagnostics Menu
Tests Menu
Screen Test
Performs a screen test by drawing lines and images on the screen
so that the operator can determine if there are any faults. Press a
keypad key to step through the test. The up and down arrow keys
will adjust the contrast.
Digital Input Test Menu
Refer to Section 3.4.18.
Digital Output Test Menu
Refer to Section 3.4.19.
3.4.18
Digital Input Test Menu
Main Menu
Service Menu
Diagnostics Menu
Tests Menu
Digital Input Test Menu
Input 0..7
Displays the status of the 0-7 digital input pins. Value will be 0 or 1.
Note:
Entering either the Digital Inputs Menu will temporarily disable all digital and analog
input/outputs. This will affect logging via these outputs. Exiting the menu restores automatic control.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...