Installation
Page 31
2.3
Instrument Set-Up
After installing the instrument the following procedures should be followed to ready the instrument
for monitoring:
Figure 5 – Instrument Back Panel
2.3.1
Pneumatic Connections
The Serinus 60 has three pneumatic ports on the back panel of the instrument: the
Sample Port
, the
Calibration Port
and the
Exhaust Port
. All tubing and fittings used should follow the instructions
below:
Must be made of Teflon® FEP material, Kynar®, stainless steel, glass or any other suitably inert
material.
Sample line should be no more than two meters in length with ⅛ inch ID and ¼ inch OD.
Sample inlet pressure should not exceed 5 kPa above ambient pressure.
Tubing must be cut squarely and any burrs removed.
Remove the inlet port nut and insert the tubing through the back of the nut with the tube
extending one inch through the front.
Place the tubing into the port until it hits the tube stop inside the fitting.
Place the nut back onto the fitting and tighten clockwise until finger tight.
Nuts should be re-tightened when instrument reaches operating temperature.
Sample Port
The sample port must be connected to an ambient source of sample air. When using a sample
manifold the Serinus requires at least 1 slpm delivered to the sample manifold (0.5 slpm for
measurement plus approximately 50 % excess).
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...