+Analog Supply
Pass if the analog power supply is within the acceptable range
(+12 V).
-Analog Supply
Pass if the analog power supply is within the acceptable range
(-12 V).
A2D
Fail only if a problem is detected with the analog to digital
conversion.
Cell Temp.
Pass if the cell heater temperature is within ±10 % of the heater
set point (CELL) (refer to Section 3.4.5).
Signal
Fail if the detector signal voltage falls below a calculated level.
Interface PCA
Pass if the Interface PCA is functioning correctly.
System Power
Pass if the system has an adequate electrical supply.
Maintenance Mode
Error if the system is “
In Maintenance
” (refer to Section 3.4.12).
Diagnostic PTF Comp
Error if the
Pres/Temp/Flow Comp.
is disabled (refer to Section
Diagnostic Control
Error if the control loop is disabled (refer to Section 3.4.13).
Valve Manual Control
Error if the valve sequencing is disabled (refer to Section 3.4.16).
Pressure Calibration
Error if the user is performing a pressure calibration.
Flow Calibration
[Internal Pump Option]
Error if the user is performing a flow calibration.
Flow Fault
Ok when the instrument has acceptable sample flow based on the
difference between cell and ambient pressures.
Flow Block Temp.
[Internal Pump Option]
Pass if the flow block heater temperature is within ±10 % of the
heater set point (FLOW) (refer to Section 3.4.5).
Chassis Temp.
Pass if the chassis temperature is within the acceptable limits
(0-50 °C).
USB Stick Disconnected
Detects whether a USB memory stick is plugged into the front USB
port.
Instrument Warmup
Ok once the instrument is out of warm-up status.
High Background
Pass if the background value is below 1100 Mm-1.
Backgrounds Disabled
Ok if the menu item
Backgrounds
in the measurement settings
menu is enabled.
3.4.5
Temperature Menu
Main Menu
Analyser State Menu
Temperature Menu
Temperature Units
The current temperature units of the instrument (
Celsius
,
Fahrenheit
or
Kelvin
).
Set Point (CELL)
The temperature set point of the measurement cell. The factory
default is 50 °C.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...