Troubleshooting
Page 115
7.
Troubleshooting
Table 7 – Troubleshoot List
Error Message/Problem
Cause
Solution
Leak in the instrument
Multiple possibilities
Refer to Section 6.3.6
Flow fault
Multiple possibilities
Refer to Section 7.2
Flow Block Temp. (Flow
block temperature
failure)
Faulty heater or
temperature sensor
Refer to Section 7.4
Cell Temp. (Measurement
cell temperature failure)
Faulty heater or
temperature sensor
Refer to Section 7.5
Noisy/unstable readings
Multiple possibilities
Refer to Section 7.3
Calibration system error
Ensure calibration system is functioning correctly
and is leak free.
Ensure sufficient gas is available for the
instrument and an adequate vent is available for
excess gas.
Leaks
A leak in the instrument or calibration system dilutes
the sample stream and causes low span readings and
noise.
Refer to Section 6.3.6
Measurement cell heater
A failed temperature control allows the instrument
to drift with ambient temperature. Verify that the
cell temperature is 50
C
3
C.
Hardware fault
Dust on mirrors of contamination in cell.
Gain too high
Leak check (repair any leaks). Refer to Section
6.3.6.
Service instrument.
Replace filters.
Instrument resetting
Multiple possibilities
Check that the instrument is not overheating.
Possibly a faulty power supply.
Corrupted firmware. Perform ‘erase all settings’ in
the
Bootloader Menu
and reload or upgrade
firmware.
12 Voltage supply failure Power supply has failed
Replace power supply.
No display
AC power
Verify that the mains power cable is connected and
the rear fan is operating.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...