
Serinus 60 User Manual 3.0
Page 106
6.3
Maintenance Procedures
6.3.1
Background Scrubber DAU check
The internal background scrubber Disposable Absorbent Unit (DAU) removes pollutants from the
sample stream providing a source of continuous NO
2
free air for background and internal zero
functions. Failure of the absorbent can result in drift, poor sensitivity near zero and high background
readings.
Note:
Ensure that the instrument has been running for at least one hour before this procedure is
performed.
Equipment Required
Span gas source 400ppb NO
2
5/8” Spanner
Procedure
1.
Open -
Main Menu
Calibration Menu
.
2.
Select -
Cal. Type
Manual
- Accept.
3.
Select -
Zero Source
Internal
- Accept.
4.
Select -
Cal. Mode
Zero
- Accept.
5.
Allow the instrument to sample Zero air for approximately five minutes.
6.
Record the instrument response as the “Initial” value.
Note:
This value should be 0.00
1 ppb. If it is not, a background should be initiated by selecting
Manual Background
from within the
Calibration menu
. The background will take approximately five
minutes to complete (refer to Section 3.4.10).
7.
Connect 400 ppb NO
2
span source to the Sample Port ensuring the inlet pressure is maintained at
ambient (vented).
8.
Allow the instrument to sample the span source for five minutes.
9.
Record the instrument response as the “Challenge” value.
10.
Disconnect the span source.
11.
Open -
Main Menu
Calibration Menu
.
12.
Select -
Cal. Mode
Measure
- Accept.
13.
Select -
Zero Source
External
- Accept.
14.
Compare the “Initial” value and the “Challenge” value. They should agree within
2 ppb
otherwise replace the background scrubber DAU.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...