Calibration
Page 93
6.
Allow the instrument time to achieve a stable response.
7.
Enter -
Zero Calibrate NO2
- OK.
8.
Disconnect the zero source and reconnect the sample line to the
Sample Port
.
5.4.3
Internal Zero (Sample Port)
Equipment Required
N/A
Procedure
1.
Open -
Main Menu
Calibration Menu
.
2.
Select -
Cal. Type
Manual
- Accept.
3.
Select -
Zero Source
Internal
- Accept.
4.
Select -
Cal. Mode
Zero
- Accept.
5.
Allow the instrument time to achieve a stable response.
6.
Enter -
Zero Calibrate NO2
- OK.
7.
Select -
Cal. Mode
Measure
- Accept (To return to sample measure).
5.5
Span Calibration
A span calibration is a calibration performed at the upper end of the instrument’s measurement
range. Ecotech recommends calibration at 80 % of the full scale measurement or operating range of
the instrument.
While the instrument range is commonly set as a default 0-500ppb, this is widely recognised as no
longer being valid with modern digital communication and most regulators will now recommend a
range more suited to local conditions.
Note:
Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure the instrument’s stability.
A span gas can be supplied through either the
Calibration Port
or
Sample Port
. Follow the relevant
instructions below:
5.5.1
Calibration Port
Equipment Required
Span source
Procedure
1.
Ensure a suitable span source is connected to the
Calibration Port
.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...