Service
Page 111
12.
Toggle Off -
Internal Pump
Off
.
13.
Note the value on the barometer. Wait for three minutes, the value should not drop more than 5
kpa (37.5 torr). If the leak check passed skip to step 16.
14.
Inspect the instrument’s plumbing looking for obvious damage. Check the condition of fittings
and sample filter holder seating plus condition of the O-ring.
15.
If the leak is still present divide up the pneumatic system into discrete sections to locate the leak
(refer to Section 7). When the location of the leak has been determined repair and then rerun the
leak check procedure.
16.
Slowly remove the barometer from the
Sample Port
and allow the instrument to return to
ambient. Inspect the internal tubing to ensuring that the tubing is cleanly connected to the
fittings and that the internal Teflon lining has not been kinked or crumpled.
17.
Open -
Main Menu
Service Menu
Diagnostics Menu
Valve Menu
.
18.
Enable -
Valve Sequencing
Enabled
.
19.
Open -
Main Menu
Service Menu
Diagnostics Menu
Internal Pump Menu
.
20.
Select -
Flow Control
Start
- Accept.
6.3.7
Clean Pneumatics
The calibration valve manifold will require disassembling and cleaning. Ideally the valves and the
manifold should be cleaned in a sonic bath with lab detergent and water. Once clean rinse with
distilled water and dry before reassembling (refer to Section 9.8 and Section 9.9). Leak test the
assembly before adding back into the system. Once the system has been reassembled the instrument
should have a full system leak check (refer to Section 6.3.6).
If the tubing shows signs of significant contamination, it should be replaced with new tubing.
6.3.8
Pressure Sensor Check
Pressure checks are needed to ensure that the pressure sensor is accurately measuring pressure
inside the instrument.
During normal operation ensure that the
Pressure & Flow Menu
indicate the following parameters.
Ambient should display the current ambient pressure at site. Cell should indicate current
measurement cell pressure depending on the pump condition and location. A Cell pressure of
between 10 torr and 50 torr is recommended.
Equipment Required
Barometer
Digital Multimeter (DMM)
Procedure
1.
Open -
Main Menu
Analyser State Menu
Pressure & Flow Menu
.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...