background image

 Electrical 

connection

 

D184B140U02 FSM4000 

45 

 

Pos: 29.29 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Elektrischer Anschluss/Anschlussbeispiele (BA) @ 17\mod_1200932005063_3101.doc @ 153602 

 

5.5 

Connection examples for peripherals (incl. HART)  

 

Current output 

 

 

G00243

+31

-32

0/4 ... 20 mA
0/2 ... 10 mA

I

E

 

I = internal 
E = external 
 
 
 
 
 

Current output: 

selectable in software 

 

Function: 

Active 
0/4 ... 20 mA (0 

 

 RB 

 560 

0/2 ... 10 mA (0 

 

 RB 

 1120 

Terminals: 31, 32 
 

(with HART only 4 ... 20 mA) 

Fig. 40:  Current output active with/without HART protocol (4 ... 20 mA) 
 
 

Contact output 

 

 

R

B

*

42

+U

G00244

41

R

B

*

U

CE

I

CE

I

E

0 V
24 V DC

 

Note 

Horn and alarm light are shown only 
as examples. Other suitable devices 
such as bells, sirens, buzzers, etc., 
can also be used. 

 
 
 

 

 

Contact output: 

selectable in software 

 

Function: 

Passive 
"closed" 0 V 

 U

CEL

 

 2 V, 2 mA 

 I

CEL

 

 220 mA 

"open" 16 V 

 U

CEH

 

 30 V, 0 mA 

 I

CEH

 

 2 mA 

Terminals: 41, 42 
 

 

Fig. 41:  Contact output for system monitoring, Max.-Min. alarm for empty tube or forward/reverse signal 
 
 

Contact input 

 

 

G00245

+ 16 ... 30 V DC

0 V

82

81

R

i

I

E

 

 

 

 

Contact input: 

selectable in software 

 

Function: 

Passive 
"On" 16 V 

 U

KL

 

 30 V 

"Off" 0 V 

 U

KL

 

 2 V 

Ri = 2 k

 

Terminals: 81, 82 
 

 

Fig. 42:  Contact input for external totalizer reset and external zero return 
 

Содержание FSM4000 Series

Страница 1: ...er FSM4000 Pos 1 Titelbl tter Copyright BA IA Durchfluss FSM4000 13 mod_1193131920109_3101 doc 131425 Valid for software versions C 10 and higher Model FSM4000 SE41F FSM4000 SE21 FSM4000 SE21F FSM4000...

Страница 2: ...y Tel 49 551 905 534 Fax 49 551 905 555 CCC support deapr de abb com Copyright 2008 by ABB Automation Products GmbH Subject to change without notice This document is protected by copyright It assists...

Страница 3: ...stallation safety information 15 1 15 Operating safety information 15 1 16 Maintenance and inspection safety information 16 2 Design and function 17 2 1 Measuring principle 17 2 2 Design 18 2 3 Device...

Страница 4: ...s 32 4 8 8 Ground with conductive PTFE grounding plate 32 5 Electrical connection 33 5 1 Preparing and routing the signal and magnet coil cable 33 5 2 Connecting the flowmeter sensor 35 5 2 1 Connecti...

Страница 5: ...essages during operation 99 9 Maintenance Repair 101 9 1 Flowmeter sensor 101 9 2 Cleaning 101 9 3 Gaskets 101 9 4 Replacing the transmitter 102 10 Spare parts list 104 10 1 Fuses for transmitter elec...

Страница 6: ...Contents 6 FSM4000 D184B140U02 13 2 Permits and certifications 118 13 3 Overview of setting parameters and technical design 122 14 Index 124...

Страница 7: ...or procedural instructions with specific safety information Only the observance of all safety information enables the optimal protection of personnel as well as the environment from hazards and the sa...

Страница 8: ...xcluded from this are repairs performed by ABB authorized specialist shops Pos 6 7 Sicherheit Durchfluss Allgemein Technische Grenzwerte Durchfluss 0 mod_1129714966562_3101 doc 3251 1 4 Technical limi...

Страница 9: ...r of liability from any resulting damages The manufacturer s warranty expires Pos 6 11 berschriften 1 1 1 spaltig S U Schilder und Symbole 0 mod_1129721947139_3101 doc 3204 1 7 Labels and symbols Pos...

Страница 10: ...SM4000 13 mod_1193139084015_3101 doc 131524 1 7 2 1 Identifying the device design 1 Identifying the model The model number of the flowmeter sensor or transmitter see nos 1 or 2 in the description of t...

Страница 11: ...s 9 Power consumption 10 Phase relationship between signal and reference voltages 11 Reference voltage 12 Cs calibration factor span 13 Cz calibration factor zero point Transmitter ABB Automation Prod...

Страница 12: ...ation of fluid group as per the pressure equipment directive PED Fluid group 1 hazardous liquids gaseous 6 Manufacturer of the pressure equipment Pressure equipment outside the applicable range of the...

Страница 13: ...Pos 6 24 Sicherheit Allgemein Organisatorische Ma nahmen R cksendung von Ger ten 0 mod_1129730744499_3101 doc 3248 1 10 Returning devices Use the original packaging or a suitably secure packaging for...

Страница 14: ...d supports the reuse of valuable raw materials If it is not possible to dispose of old equipment properly ABB Service can accept and dispose of returns for a fee Pos 6 28 Sicherheit Durchfluss Allgeme...

Страница 15: ...e the supplied cable D173D147U01 for the magnet coil circuit This does not apply for older model flowmeter primaries 10D1422 10DI1425 DN 500 In this event the magnet coil is supplied via the supply po...

Страница 16: ...dium which spurts out can cause severe injuries Depressurize pipes before opening the inspection screw Corrective maintenance work may only be performed by trained personnel Depressurize the device an...

Страница 17: ...etrically opposite each other This signal voltage UE is proportional to the magnetic induction B the electrode spacing D and the average flow velocity v Considering that the magnetic induction B and t...

Страница 18: ...rate location from the flowmeter sensor Up to 50 m cable length for a minimum conductivity of 20 S cm For flowmeters with preamplifiers the signal length is increased to 200 m The electrical connectio...

Страница 19: ...mod_1193319756546_3101 doc 132475 3 3 Transport of flanged units smaller than DN 350 Warning Danger of injuries due to slipping meter The center of gravity for the complete device may be higher than...

Страница 20: ...Transport 20 FSM4000 D184B140U02 G00461 Fig 9 Transport of flanged units larger than DN 300 Pos 16 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...

Страница 21: ...r torques on the device Do not remove the plugs in the cable connectors until you are ready to install the electrical cable Make sure the gaskets for the housing cover are seated properly Carefully se...

Страница 22: ...ocks can destroy the meter 1 Remove protective plates if present to the right and left of the measuring tube To prevent possible leakage make sure that the liner on the flange is not cut or damaged 2...

Страница 23: ...10 25 1 CL 300 4 x M16 18 40 4 x M16 34 CL 150 4 x M12 15 32 1 1 4 CL 300 4 x M16 27 40 4 x M16 43 CL 150 4 x M12 20 40 1 1 2 CL 300 4 x M20 43 40 4 x M16 56 CL 150 4 x M16 39 50 2 CL 300 8 x M16 28...

Страница 24: ...M20 115 16 12 x M24 160 25 16 x M27 175 40 16 x M30 340 CL 150 12 x M24 180 300 12 CL 300 16 x M30 265 10 16 x M20 145 16 16 x M24 195 350 14 25 16 x M30 280 10 16 x M24 200 16 16 x M27 250 400 16 25...

Страница 25: ...4 x M16 26 CL 150 4 x M12 upon request 32 1 1 4 CL 300 4 x M20 upon request 40 4 x M16 33 CL 150 4 x M12 upon request 40 1 1 2 CL 300 4 x M20 upon request 40 4 x M16 46 CL 150 4 x M16 upon request 50...

Страница 26: ...t t 17 mod_1201520634531_3101 doc 154368 The device may not be installed vertically with the terminal box or transmitter housing pointing downward G00528 Fig 12 The bracket mounting option no longer a...

Страница 27: ...e outlet pipe section 2 For compliance with the measuring accuracy observe the inlet and outlet pipe sections G00037 1 2 3xDN 2xDN Fig 17 Pos 21 6 Spaltenumbruch 0 mod_1132937966324_3101 doc 3831 Pos...

Страница 28: ...ennweiten 3 mod_1157632081771_3101 doc 39865 4 6 8 Installation in larger size pipelines Determine the resulting pressure loss when using reduction pieces 1 1 Calculate the diameter ratio d D 2 Determ...

Страница 29: ...1 2 3 40 40 40 0 0 0 3 6 9 m 3 h m 3 h m 3 h 13 26 40 US gal min US gal min US gal min 0 0 0 60 120 180 m 3 h m 3 h m 3 h 264 528 793 US gal min US gal min US gal min 100 125 150 4 5 6 16 16 16 0 0 0...

Страница 30: ...ground via the terminals is not required Important If the flowmeter sensor is installed in plastic or earthenware pipelines or in pipelines with an insulating lining transient current may flow through...

Страница 31: ...tallische Rohre bzw Rohre mit isolierender Auskleidung Standard 0 mod_1133359180095_3101 doc 3365 4 8 4 Non metallic pipes or pipes with insulating liner For plastic pipes or pipes with insulating lin...

Страница 32: ...bei Ger ten mit Schutzscheiben Ex und Standard 0 mod_1133281076517_3101 doc 3358 4 8 7 Ground for devices with protective plates The protective plates are used to protect the edges of the liner in th...

Страница 33: ...length and terminate both cables as shown Important Use wire end sleeves Signal cable D173D025U01 Magnet coil cable D173D147U01 Preparing the transmitter side G00535 15 5 30 1 18 8 8 80 3 15 60 2 36 0...

Страница 34: ...create stray fields switching pulses and induction If this is not possible run the signal magnet coil cable through a metal pipe and connect this to the station ground All leads must be shielded and...

Страница 35: ...nsmitter over terminals M1 M3 Connect the cables to the flowmeter sensor according to the following drawing using a screwdriver with proper size and width G00553 SE SE 3A 3A 1S 2S 1 2 3 16 16 M1 M3 1...

Страница 36: ...nnection 1 Use the supplied cable to connect the flowmeter sensor and the transmitter 2 Connect the signal cable in the terminal box of the flowmeter sensor 3 Route the cable from the terminal box to...

Страница 37: ...installation before beginning sealing activities in order avoid moisture penetration Before starting check all the connections for correct fitting and stability Do not overfill the terminal box Keep...

Страница 38: ...s Durchfluss FSM4000 Einbau der Hochtemperatur Ausf hrung 15 mod_1196072454125_3101 doc 142727 When installing the high temperature design with a fluid temperature of max 180 C the connection box for...

Страница 39: ...to terminal L phase N neutral or 1 2 and as stated on the name plate The supply power connecting cable must be rated for the current consumption of the flowmeter system The cables must comply with IE...

Страница 40: ...r and transmitter according to the connection diagram G00555 FRAM 3 1 2 Fig 35 1 Signal cable 2 Magnet coil cable 3 Busbar SE Important The power supply of the FSM4000 with preamplifier is connected v...

Страница 41: ...0 DN 1000 3 8 40 1 Functional ground busbar 2 Pulse output see page 46 3 Contact input see page 45 4 Contact output see page 45 5 Current output see page 45 6 Magnet coil cable shielded 2 x 1 mm2 CE T...

Страница 42: ...elded 2 x 1 mm2 CE type 227 TEC 74 ABB order no D173D147U01 10 m included in shipment standard 7 Supply power Potent voltage 100 230 V AC terminals L N Low voltage 20 4 26 4 V AC 20 4 31 2 V DC Termin...

Страница 43: ...mount housing 1 Functional ground busbar 2 Pulse output see page 46 3 Contact input see page 45 4 Contact output see page 45 5 Current output see page 45 6 Supply power Potent voltage 100 230 V AC ter...

Страница 44: ...F FF 3a Connection example via M12 plug see page 47 3c Bus termination with installed bus termination 3a components with hook switches closed 4 Magnet coil cable shielded 2 x 1 mm2 CE Typ 227 TEC 74 A...

Страница 45: ...ART protocol 4 20 mA Contact output RB 42 U G00244 41 RB UCE ICE I E 0 V 24 V DC Note Horn and alarm light are shown only as examples Other suitable devices such as bells sirens buzzers etc can also b...

Страница 46: ...ulse width 50 ms max pulse frequency 3 Hz 500 Load 10 k Pulse width 0 1 ms max pulse frequency 5 kHz Passive closed 0 V UCEL 2 V 2 mA ICEL 220 mA open 16 V UCEH 30 V 0 mA ICEH 2 mA Terminals 51 52 Fig...

Страница 47: ...S U 3 3 2 16 6 7 Connection board for field housing unit Hook switches for bus termination must be closed Fig 45 Connection via M12 plug for PROFIBUS PA only As an option the bus can also be connected...

Страница 48: ...must be observed When flowmeter sensor SE41F SE21 SE21F and transmitter S4 are delivered as a pair the data memory module external FRAM is located in the flowmeter sensor The data module stores the f...

Страница 49: ...the end numbers Y1 Y2 etc End numbers X1 and Y1 are considered a unit Check of pulse output The pulse output can be operated as active output 24 VDC pulse or as passive output optocoupler The current...

Страница 50: ...r to the Entering data in short form A short overview of the menu structure can be found in the section Parameter overview The Easy Set up menu enables users to configure the unit quickly and convenie...

Страница 51: ...r When the DEP mode Standard is selected it is not required to perform an adjustment onsite The transmitter works with standard setup data If the function is not performed correctly a new adjustment m...

Страница 52: ...en the Program protection is turned off No STEP Jump to menu level Yes ENTER Language English Unit Qmax l s Unit totalizer m3 Qmax 50 000 l s Damping 1 Tau 0 22 sek Display mode 4 small Display 1st li...

Страница 53: ...appear once in the segment The address can be set either locally on the unit via the DIP switches on the digital board using system tools or via a PROFIBUS DP master class 2 such as SMART VISION DSV4...

Страница 54: ...Switches 1 5 7 ON means 1 16 64 81 Bus address 81 Switch 1 2 3 4 5 6 7 8 9 A Status Unit address Address mode no function no function Off 0 0 0 0 0 0 0 Bus no function no function On 1 2 4 8 16 32 64...

Страница 55: ...onsumption Pos 31 14 Inbetriebnahme Durchfluss FSM4000 Durchf hrung der Inbetriebnahme Hinweise zur Spannungs Stromaufnahme 16 mod_1196244930750_3101 doc 143177 Da sEinschaltverhalten entspricht dem E...

Страница 56: ...Durchf hrung der Inbetriebnahme Inbetriebnahme von FOUNDATION FIELDBUS Ger ten 16 mod_1196246119343_3101 doc 143335 For units with a FOUNDATION fieldbus the settings of the DIP switch must be checked...

Страница 57: ...ss is automatically allocated at the FF via LAS link active scheduler For address detection a unique number is used DEVICE_ID This number is a combination of manufacturer ID device ID and device seria...

Страница 58: ...Line Totalizer value forward R 516 0000 m3 3 Line Totalizer value reverse multiplex mode Totalizer overflow always occurs at a value of 9 999 999 units If the totalizer value for a flow direction is l...

Страница 59: ...se outputs show the current operating mode The functions of the individual keys are explained in the following C CE Toggle between operating mode and menu STEP The STEP key is one of two arrow keys Us...

Страница 60: ...lay flashes to indicate that the function is active There are two different methods of entering data Numeric entry Entry from specified table Important When entering data the values are checked for pl...

Страница 61: ...on Qmax Submenu 5 Find parameter Qmax STEP or DATA 1800 00 m h 5 Parameter Submenu unit STEP or DATA Unit Qmax Unit Qmax 6 Change parameter Qmax ENTER m h 6 Change parameter m h ENTER m h Min 500 m h...

Страница 62: ...the menu is locked and requires the correct PP code To suspend the PP Code function you must access the PP Code menu Select ENTER and you will be asked for the old PP code You must enter the correct o...

Страница 63: ...andard Standard inverse reverse the flow direction indicators in the display for forward and reverse flow Noise reduction Off Off 1 2 3 4 The noise reduction is switched on when the measured value is...

Страница 64: ...on liters default value in kgal units factory setting Prog Unit without density Without density With density Unit for mass with density or volume without density Density 1 g cm3 Numeric entry Density...

Страница 65: ...ected Submenu Data 70 Hz This menu does not appear with an excitation frequency of 50 Hz or 60 Hz Channel 123 s View only Adjustment value for Channel can be monitored for the excitation frequency set...

Страница 66: ...menu is displayed at initial startup Additional settings can be entered in the system data submenu Primary See chapter 11 Frequency 50 Hz View only Display the mains frequency for which the flowmeter...

Страница 67: ...ENTER and STEP to access the parameters Mains Frequency Type of Sensor Meter Size and Reference Voltage Not displayed if new sensor SE41F SE21 is selected See chapter 11 Frequency 50 60 Hz 50 Hz Displ...

Страница 68: ...f the fluid level drops below the electrode level all outputs can be automatically set to zero This ensures false pulses or incorrect displays are avoided when the meter pipe is empty In addition to t...

Страница 69: ...justment values full and empty are determined and the threshold based on these values is optimized the transmitter always uses only the most recent adjustment value The threshold value is the basic cr...

Страница 70: ...alue for Volume_Flow Transducer block Index 17 1st line TB VolFlow Status TB VolFlow Status Displays the status of Volume_Flow Transducer block Index 17 1st line TBT Value 1st line TBT Status TB Total...

Страница 71: ...talizer R Totalizer diff Detector empty pipe additional options FF address displayed in hexadecimal 1st line TB VolFlowValue Displays the value or status of the flowrate TB Primary_Value Index 14 1st...

Страница 72: ...figured flow range Configurable in 1 increments 1 switching hysteresis Submenu Totalizer The totalizer function is set automatically in the Operating mode submenu Standard operating mode standard tota...

Страница 73: ...current output range For HART protocol always 4 20 mA Iout at Alarm Low Alarm Low Alarm High Alarm Behavior of the current output during an alarm Low Alarm is available only for current output of 2 10...

Страница 74: ...internal totalizer and overflow values when a switching voltage is connected For Ext Zero Return all outputs are switched off if a switching voltage is connected Example Undefined filled pipeline cond...

Страница 75: ...ilable for standard devices without HART hardware Instr address 0 Numeric entry 0 15 For devices with HART protocol an address between 0 15 can be selected If the address entered is 0 the current outp...

Страница 76: ...ays the settings for the DIP switches OFF X ON If DIP switch 8 ON The address is binary coded using switches 1 7 Switches and 9 and A have no function Example Set address 50 using the switches 50dec 3...

Страница 77: ...ce in the data link submenu It can also be read via the FF communication in the transducer block An open switch is shown as a closed switch as x The following switches are assigned Switch 1 Simulate E...

Страница 78: ...0 mA 2 10 mA This menu does not appear for the PROFIBUS PA or FOUNDATION fieldbus Test the current output Check the value displayed at the connection terminals using a digital voltmeter or the process...

Страница 79: ...tervall Cycle Time Manual Manual 10 s 60 s 10 min 60 min 6 h 12 h 24 h 7 d Select the time interval between the individual diagnostic measurements For the values for Coil Resistance Coil Temperature I...

Страница 80: ...or alarm 2 Ohm Max Wert 0 500 Upper range limit for alarm 500 Ohm Status signal Defines the alarm behavior Off Off warning error Note If coil resistance is being measured at least every 60 s the last...

Страница 81: ...lectrode AC voltage for electrode E1 Min Val 0 3000 Lower range limit for alarm 0 0 V Max Val 0 3000 Upper range limit for alarm 3000 0 V Status signal Defines the alarm behavior Off Off warning error...

Страница 82: ...operly display the electrode voltage the electrode zero must be adjusted For additional information see Submenu Adjustment on page 85 Submenu Alert Mag Lin Set the thresholds and alarm behavior for th...

Страница 83: ...alarms to the default values recapture Yes exit via C CE Yes Enter Submenu Conn Error Submenu CoilMassShCut Defines alarm behavior in case of an accidental coil ground Note If the coil circuit is bei...

Страница 84: ...Magnetic linearity SNR Signal to noise ratio Max DAC value Max DAC value Oprtg hrs trn Operating hours for transmitter Oprtg hrs trn Operating hours for flowmeter sensor These values are updated only...

Страница 85: ...rence resistance for the coil See also chapter 7 4 Additional information regarding use of enhanced diagnostic functions Note This menu is displayed only when the coil resistance or coil temperature a...

Страница 86: ...s Warning register Number 2 3 Pending Test Mode Reset Enter Help text Enter 2 Totalizer reset After calling up the warning register the warning register text is displayed with the number of pending wa...

Страница 87: ...must be at absolute standstill Press ENTER to perform the automatic adjustment The limit for the zero point is 10 If the value is outside this limit no calibration is performed FSM4000 01 2008 Displa...

Страница 88: ...estimmung der Diagnose Messwerte 16 mod_1196845404828_3101 doc 144420 7 4 1 1 Determining the AC coil current or coil AC voltage I U G00578a Detecting changes in the coil circuit The current and volta...

Страница 89: ...s SE41F SE21 and S4 When connecting an S4 transmitter to already shipped sensor SE41F SE21 it is necessary to perform the adjustment For additional information see chapter 7 4 3 Readjusting the coil t...

Страница 90: ...85 7 4 1 6 Determining the electrode balance E1 E2 EB 2 E 1 E E1 E2 EB 1 E 2 E Detect distorted flow profile e g due to improper installation Detect interference in electrode circuit e g failure of el...

Страница 91: ...ics Alert Settings SNR sensor flowmeter from Signal signal Filtered 7 4 1 8 Determining the magnetic linearity I UR G00578f ML 1 R 2 2 R 1 U I U I Detect magnetic field interference This can affect th...

Страница 92: ...d value can be displayed To differentiate between the values in the display an identification code ID is prefixed before to each value As a result each line appears as follows e g CV_RCL 80 5 ohm for...

Страница 93: ...t flowmeter sensor log time H Min Sec FVP Factory Fingerprint flowmeter sensor log time H Min Sec H1P H9P History 1 9 flowmeter sensor log time H Min Sec 7 4 2 2 Diagnostic warning and error messages...

Страница 94: ...in operation EMx Curr Max Errors Error Max present available current overshoots EMi Curr Min Errors Error Min present available current undershoots WMx Curr Max Warn Warning Max present available curr...

Страница 95: ...196860571562_3101 doc 144633 7 5 Software history 7 5 1 For transmitters without communication or HART protocol Software D200S021U01 Software Version Type of changes Documentation supplements B 10 OEM...

Страница 96: ...und Alarmierungen 16 mod_1196864174421_3101 doc 144680 8 1 Overview of error states and alarms Contact output State Error Flowrate indicator display Notification with simulation at current output Cur...

Страница 97: ...he AD converter DSP is defective Replace the DSP board Error 2 Driver 7 Positive or negative reference too small Check the cables no reference voltage is present The current limiter in the driver has...

Страница 98: ...he operating mode submenu Use a air separator clean electrodes activate empty pipe detector Contact ABB Service Error F FRAM in Primary 1 Missing data from external FRAM FRAM for the flowmeter sensor...

Страница 99: ...outine Warning 71 Ext Dat loaded 9 Transmitter has recognized different flowmeter sensor data and loaded it in the internal FRAM Loaded are the system and flowmeter sensor data Defective FRAM repair u...

Страница 100: ...se output on Exit the functional test of the pulse output Warning 132 3 Auto Adjust 15 Adjustment started for system zero point adjustment via the contact input for internal adjustment for ABB Service...

Страница 101: ...eter sensor electrodes must be cleaned when the flowrate information on the transmitter changes when recording the identical flowrate volume If the display shows a higher flowrate the contamination is...

Страница 102: ...ldbus If the transmitter is located at the bus end and if the bus termination is activated via both hook switches the bus termination is canceled when the transmitter plug in module is removed The bus...

Страница 103: ...ess data is imported from the internal to external FRAM The flowmeter sensor data continues to be imported from the external to internal FRAM The following message is displayed Warning 8b Update exter...

Страница 104: ...l connecting cables must be switched off Replacing the fuse 1 Switch off supply power After waiting 40 minutes remove the housing cover from the field housing unit 2 To replace the fuses for the remov...

Страница 105: ...n board standard Connection board PA FF D685A1020U03 D685A1020U02 3 Cover large complete D641A030U01 4 PVC cover D626A005U01 5 Spacer bolts M4 x 75 galvanized steel D124E009U20 6 Cover small D641A029U...

Страница 106: ...er number 1 Connection board standard DN 8 D685A1025U01 1 1 Connection board with preamplifier D685A1028U01 2 Lower section SS 1 4301 model SE21_ D612A128U01 3 Gasket D333F016U01 4 Pan head screw M3 x...

Страница 107: ...DN 100 D612A153U03 2 3 Lower section with NPT adapter DN 100 D612A153U07 3 Cheese head screw M3 x 8 DIN 84 D002F107AU20 3 1 Clamp strap D108A003U08 4 Sheet metal screw 2 9x6 5 DIN 7981 D055E106CZ01 4...

Страница 108: ...nce voltage Meter factor Entries can be made one after another In the submenu Primary Setup select the mains frequency at which the flowmeter sensor was calibrated See name plate Select the flowmeter...

Страница 109: ...itter S4 from name plate 10DS3111 A C 10DS3111 D 10DS3111 E 500 500 500 no Cs calculated Cz 0 External supply power from name plate 10D1422 10DI1425 3 1000 500 yes yes Cs calculated Cz 0 External supp...

Страница 110: ...ting cables must be switched off Adapting the transmitter to the flowmeter sensor 10D1422 DN 3 1000 10DI1425 DN 500 1 Switch off supply power After waiting 40 minutes remove the housing cover from the...

Страница 111: ...24106281_3101 doc 140273 12 2 Flowmeter sensor Pos 48 6 berschriften 1 1 1 2 spaltig Messwertaufnehmer SE41F 17 mod_1200930203802_3101 doc 153533 12 2 1 Flowmeter sensor SE41F Pos 48 7 Technische Date...

Страница 112: ...k frame dark gray RAL 7012 Cover light gray RAL 9002 Meter tube SS 1 4301 304 PG Connector Polyamide Storage Temperature 20 70 C 4 158 F Protection Class per EN 60529 IP 67 IP 68 option Pipeline Vibra...

Страница 113: ...5 290 0 362 5 435 0 507 5 580 0 652 5 725 5 797 5 0 Fig 62 ASME flange SS 1 4571 316 Ti to DN 300 CL150 300 to DN 1000 CL150 G00217 PS bar TS 30 10 10 30 50 70 90 110 130 150 170 190 C 22 14 50 86 12...

Страница 114: ...101 doc 153556 12 2 2 Flowmeter sensor SE21 SE21F Pos 48 10 Technische Daten Datenblatt Durchfluss FSM4000 Technische Daten Messaufnehmer Modell SE21_ 3 mod_1157640761823_3101 doc 40390 Minimum Allowa...

Страница 115: ...tion Tri Clamp External threads EPDM Ethylene Propylene with FDA approval silicone with FDA approval CIP resistant no oils or grease Silicon with FDA approval optional resistant to oils and grease PTF...

Страница 116: ...ndard 20 S cm with preamplifier 5 S cm with preamplifier 0 5 S cm with preamplifier DN 10 1000 3 8 40 DN 1 8 1 25 5 16 DN 1 1000 1 25 40 DN 10 1000 3 8 40 Reproducibility measurement period 100 s DN 1...

Страница 117: ...rgeh use und Montagevorschlag 0 mod_1131704576853_3101 doc 3671 12 4 1 Transmitter housing and mounting arrangements G00073 249 9 80 167 6 57 7 0 27 38 1 49 83 5 3 28 139 7 5 50 66 2 59 265 10 43 14 5...

Страница 118: ...h the following directives and their basic safety requirements CE mark on the name plate of transmitter Conforms with EMC directive 89 336 EWG Conforms with low voltage directive 73 23 EWG CE mark on...

Страница 119: ...Appendix D184B140U02 FSM4000 119 Pos 62 Zertifikate Durchfluss FSM4000 EG Konformit tserkl rung 18 mod_1203435138428_0 doc 163561...

Страница 120: ...Appendix 120 FSM4000 D184B140U02...

Страница 121: ...Appendix D184B140U02 FSM4000 121 Pos 63 Seitenumbruch 0 mod_1126532365768_3101 doc 3830...

Страница 122: ...alarm Empty pipe etc Contact input External zero return Totalizer reset External system zero no function Current output 0 4 20 mA 0 2 10 mA 0 5 mA 0 10 20 mA 4 12 20 mA Iout with alarm 0 103 3 8 mA L...

Страница 123: ...son Telephone Fax E Mail Device details Type Serial no Reason for the return description of the defect Was this device used for working with substances which pose a threat or health risk F Yes F No If...

Страница 124: ...Factory Tag 12 Field housing unit 105 Flowmeter sensor 101 111 Free inlet or outlet 27 G Gaskets 22 101 General information on ground connections 30 General information on transport 19 General Safety...

Страница 125: ...meter sizes larger than DN 300 21 Symbols and warnings 9 System integration 55 T Technical data 111 Technical limit values 8 Temperaturdiagramm 111 Terminal connection diagrams 41 Torque information...

Страница 126: ...without notice Printed in the Fed Rep of Germany 06 2008 ABB 2008 D184B140U02 Rev A ABB Limited Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel 44 0 1453 826661 Fax 44 0 1453 829671 ABB Inc 12...

Отзывы: