Parameterization
D184B140U02 FSM4000
85
Submenu/parameter
Value range
Comment
to Startup
Copies the current data record in the Fingerprint "as
Startup" (Startup Fingerprint).
Transfer
Yes
→
Enter
Yes, exit via C/CE
Submenu
Adjust
Submenu
CoilTemp
Temp. °C
0.0
-100 ... 100
Enter the current coil temperature to readjust the coil
resistance at 20 °C. Perform adjustment only on a "cold"
device. See chapter "7.4 Additional information regarding
use of enhanced diagnostic functions".
Submenu
CablLeng
CablLeng
0.0
0 ... 200
Enter the length of the signal cable.
Note:
Information must be entered to calculate the coil
resistance / coil temperature.
Submenu
Temp. Offset
Temp. °C
0.0
-100 ... 100
Offset correction for temperature display.
Transfer RT20
Transfer
New
500
Ohm
Old
450
Ohm
Yes
→
Enter
Yes, exit C/CE
Transfer the reference resistance for the coil.
See also chapter "7.4 Additional information regarding
use of enhanced diagnostic functions".
Note:
This menu is displayed only when the coil
resistance or coil temperature are enabled as an error or
warning message.
ElecZero
This adjustment is required in order to use the diagnostic
measurement values for the electrode voltage E1/E2 and
the electrode balance as well as the wiring errors for the
electrode state.
ElecZero
New 150 180 µV
Old 200 190 µV
Yes
→
Enter
Yes, exit C/CE
Display the current (new) and stored (old) zero values for
electrode E1/E2.
Select "Enter" to import the current zero points
Note:
The fluid in the flowmeter sensor must be at
absolute standstill.
ElecAnalyst
N1:
12
µV
286°
N2:
10
µV
283°
A1:
650
µV
133°
A2:
670
µV
135°
Yes, exit C/CE
Displays stored electrode zero and current electrode AC
voltage acc. to magnitude and phase.
Where
N1: Current measured electrode AC voltage E1
N2: Current measured electrode AC voltage E2
A1: Stored zero E1
A2: Stored zero E2