Technical data
114 FSM4000
D184B140U02
JIS 10K-B2210 Flange
Nominal
size DN
Material
PN
TS
PS
[bar]
32 … 300
(1¼ … 12“)
SS 1.4571
(316 Ti)
10
-25 … +180 °C
(-13 … +356 °F)
10
32 … 300
(1¼ … 12“)
Steel
10
-10 … +180 °C
(14 … 266 °F)
10
G00219
PS
[bar]
TS
6
7
8
9
10
11
12
13
14
15
16
17
DN 700 PN 16
DN 900 PN 16
DN 800 PN 16
DN 1000 PN 16
DN 900 PN10
DN 800 PN 10
DN 700 PN 10
DN 1000 PN 10
-30
-20
-10
0
10
20
30
40
50
60
70
80
90 [°C]
-22
-4
14
32
50
68
86
104
122
140
158
176
194 [°F]
PS
[psi]
87.0
101.5
116.0
130.5
145.0
159.5
174.0
188.5
203.0
217.5
232.0
246.5
Fig. 65: DIN flange SS 1.4571 (316 Ti), DN 700 ... DN 1000
(28 ... 40“)
G00220
PS
[bar]
TS
DN 700 PN 16
DN 900 PN 16
DN 800 PN 16
DN 1000 PN 16
DN 900 PN 10
DN 800 PN 10
DN 700 PN 10
DN 1000 PN 10
6
7
8
9
10
11
12
13
14
15
16
17
-10
0
10
20
30
40
50
60
70
80
90 [°C]
-14
32
50
68
86
104
122
140
158
176
194 [°F]
PS
[psi]
87.0
101.5
116.0
130.5
145.0
159.5
174.0
188.5
203.0
217.5
232.0
246.5
Fig. 66: DIN flange steel, DN 700 ... DN 1000 (28 ... 40“)
Pos: 48.8 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
Pos: 48.9 /Überschriften/1.1.1/2-spaltig/Messwertaufnehmer SE21 / SE21F @ 17\mod_1200930648191_3101.doc @ 153556
12.2.2 Flowmeter sensor SE21 / SE21F
Pos: 48.10 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messaufnehmer/Modell SE21_ @ 3\mod_1157640761823_3101.doc @ 40390
Minimum Allowable Absolute Pressure
Liner
Nominal size
DN
P
Operation
mbar abs
at
T
Operating
*
°C
PFA
3 … 100
(1/10 … 4“)
0
≤
130
°C
(266 °F)
Peek/Torlon 1
…2
(1/25 … 1/12“)
0
≤
120
°C
(248 °F)
1) Higher temperatures are allowed for CIP/SIP cleaning for limited
time periods, see Table "Maximum Allowable Cleaning
Temperature".
Maximum Allowable Cleaning Temperature
CIP-Cleaning
Liner
T
max
T
max
Minutes
T
amb.
Steam cleaning
PFA / Peek
150 °C
(302 °F)
60
25 °C
(77 °F)
Liquid cleaning
PFA / Peek /
Torlon
140 °C
(284 °F)
60 25
°C
(77 °F)
If the ambient temperature is > 25 °C (77 °F), then the difference
must be subtracted from the max. cleaning temperature.
T
max
–
Δ
°C,
Δ
°C = (T
Amb
- 25 °C)
Maximum Allowable Temperature Shock
Liner
Temp Shock max.
Temp. Diff. °C
Temp. gradient
°C/min
PFA any any
Peek, Torlon
any
any
Maximum allowable ambient temperature as function of
fluid temperature
T
amb
°C (°F)
G00224
-40
60 (140)
40 (104)
-10
(14)
-10
(14F)
DN 1 ... DN 2
DN
1
...
DN
2
90 (194)
120
(248)
130
(266)
-25 (-13)
T
Measuring
°C (°F)
Fig. 67: Temperature graph
The flowmeter sensor may not be isolated
Fluid temperature DN 1 … DN 2 (1/25 ... 1/12")
-10 … 120 °C (14 ... 248 °F), max. allowable cleaning temperature,
see table