![ABB FSM4000 Series Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/abb/fsm4000-series/fsm4000-series_operating-instruction_3770914094.webp)
Parameterization
94 FSM4000
D184B140U02
The error registers for diagnostic values are structured as follows:
ID code
_______________________________
Set Off
Description
S
−
Signal
quality
L
−
Magnetic
linearity
B
−
Electrode
balance
e
−
Electrode voltage E1
E
−
Electrode voltage E2
D
−
DAC
value
I
−
Insulation
resistance
T
−
Coil
temperature
L
−
Coil
resistance
R
−
Coil
voltage
C
−
Coil
current
↓
ID
code
Register name
Content of register
HMx
Hist Max Errors
Error / Warning:
(Max.) values overshot while transmitter is in
operation.
HMi
Hist Min Errors
Error / Warning:
(Min.) values undershot while transmitter is in
operation.
EMx
Curr Max Errors
Error:
(Max.) present available current overshoots.
EMi
Curr Min Errors
Error:
(Min.) present available current undershoots.
WMx
Curr Max Warn
Warning:
(Max.) present available current overshoots.
WMi Curr
Min
Warn
Warning:
(Min.) present available current undershoots.
The information EMx _R______B__ indicates Max. Alarm in the current values for the coil
voltage and electrode balance.
The error registers for wiring errors are structured as follows:
ID code
_______________________________
Set Off
Description
M
−
Coil short-circuit to
ground
E
−
↓
Wiring error for
electrodes
ID code
Register name
Content of register
ConWarn
Wiring warning
Current warnings for wiring
ConErr
Wiring Error
Current errors for wiring
Pos: 35.8 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830