Parameterization
80 FSM4000
D184B140U02
Submenu/parameter
Value range
Comment
Submenu
Alert Coil Curr
Set the thresholds and alarm behavior the AC coil current.
Min. Val
0 ... 500
Lower range limit for alarm.
20.0 mA
Max. Val
0 ... 500
Upper range limit for alarm.
500.0 mA
Status signal
Defines the alarm behavior.
Off
Off, warning, error
Submenu
Alert coil resist
Set the thresholds and alarm behavior the coil resistance.
Note: To correctly display the coil resistance or coil
temperature, may be an adjustment is required and the
cable length must be entered. For additional information,
see "Submenu Adjustment" on page 85.
Min. Wert
0 ... 1500
Lower range limit for alarm.
2 Ohm
Max. Wert
0 ... 500
Upper range limit for alarm.
500 Ohm
Status signal
Defines the alarm behavior.
Off
Off, warning, error
Note:
If coil resistance is being measured (at least every
60 s), the last valid flowrate value during the measure-
ment is retained (approx. 2 s).
Submenu
Alert coil temp
Set the thresholds and alarm behavior the coil
temperature.
Note: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see
"Submenu Adjustment" on page 85.
Min. Val
-100 ... 200
Lower range limit for alarm.
-50 °C
Max. Val
-100 ... 200
Upper range limit for alarm.
200 °C
Status signal
Defines the alarm behavior.
Off
Off, warning, error
Note:
If coil resistance is being measured (at least every
60 s), the last valid flowrate value during the measure-
ment is retained (approx. 2 s).