![ABB FSM4000 Series Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/abb/fsm4000-series/fsm4000-series_operating-instruction_3770914033.webp)
Electrical
connection
D184B140U02 FSM4000
33
Pos: 29.1 /Überschriften/1/D - F/Elektrischer Anschluss @ 0\mod_1140620471687_3101.doc @ 3143
5 Electrical
connection
Pos: 29.2 /Elektrischer Anschluss/Durchfluss/FSM4000/Konfektionierung des Signal- und Magnetspulenkabels @ 14\mod_1193742076406_3101.doc @ 135066
5.1
Preparing and routing the signal and magnet coil cable
Cut to length and terminate both cables as shown.
Important
Use wire end sleeves.
Signal cable D173D025U01
Magnet coil cable D173D147U01
Preparing the transmitter side
G00535
15
5
30 (1.18)
8
8
80 (3.15)
60 (2.36)
(0.20)
(0.59)
17 (0.67)
(0.31)
17 (0.67)
(0.31)
Preparing the flowmeter sensor side
G00549
5
4
3
2
1
G00552
M1
M3
1
2
3
Cable identifiers
Fig. 29: Dimensions in mm (inch)
1
Measurement potential, yellow
2
Reference, white
3
Signal line, red
4
Signal line, blue
5
SE clamp
1
Magnet coil, black
2
Magnet coil, black
3
SE clamp
Important
The shields may not touch (signal short circuit).