Startup
Operation
D184B140U02 FSM4000
49
• The flowmeter sensor must be installed at a largely vibration-free location.
• The flowmeter sensor and the converter must be assigned properly for the model FSM4000:
The flowmeters sensor have an end number of X1, X2, etc., on the model plate. The
transmitters have the end numbers Y1, Y2, etc. End numbers X1 and Y1 are considered a
unit.
• Check of pulse output.
The pulse output can be operated as active output (24 VDC pulse) or as passive output
(optocoupler). The current setting is provided on the model plate for the transmitter.
Modification as shown in the following illustration.
G00559
1
A
B
Fig. 48: Setting the pulse output using jumpers
1
Display board
A Pulse passive (jumper inner)
B Pulse active (jumper outer)