Parameterization
D184B140U02 FSM4000
69
Submenu/parameter
Value range
Comment
Adj. empty pipe
Manual
1000
Autom. Adjust
Adjustment value for empty meter pipe. The pipeline must
be empty. After the adjustment is completed, the
transmitter displays the adjustment value and calculates a
new threshold value.
Not displayed, if DEP is switched off or DEP mode is
standard.
Adj. full pipe
Manual
500
Autom. Adjust
Adjustment value for full meter pipe. The pipeline must be
filled. After the adjustment is completed, the transmitter
displays the adjustment value and calculates a new
threshold value.
Not displayed, if DEP is switched off or DEP mode is
standard.
Threshold
Numeric entry
100000
Threshold value for turning off the "Empty pipe" function.
This value is calculated and entered internally after the
adjustment "Empty pipe" or "Full pipe" is completed. It can
also be entered manually (e.g., after both adjustment
values (full and empty) are determined and the threshold,
based on these values, is optimized (the transmitter
always uses only the most recent adjustment value)). The
threshold value is the basic criteria used for the DEP
function. The adjustment values are used only to
calculate the threshold.
Not displayed, if DEP is switched off or DEP mode is
standard.
Alarm. e. pipe
Off
Off
On
If the alarm is switched on, the empty pipe contact is
actuated when the pipe empties. The "empty pipe" or
"General alarm" functions must be selected in the "Progr.
in/output" submenu. When the function is turned off, the
alarm does not respond.
Not displayed, if DEP is switched off.
Iout at e. pipe
Low Alarm
Low Alarm
High Alarm
0%
Set the current output value to Low or High in event of
error.
Not displayed, if DEP is switched off.
Submenu
Display
Select each different line displayed for the process
indicators in this submenu. You can also adapt the
contrast of the display to local conditions by pressing the
arrow keys.
Display mode
1
big,
1
small
1 big, 1 small,
4 small
Select 1 big and 1 small line or 4 small lines.