Appendix
122 FSM4000
D184B140U02
Pos: 64 /Anhang/Durchfluss/FSM4000/Übersicht Geräteparameter und technische Ausführung @ 16\mod_1197298637937_3101.doc @ 145521
13.3 Overview of setting parameters and technical design
Measuring point:
TAG no.:
Flowmeter model:
Transmitter type:
Order no.:
Device no.:
Order no.:
Measured medium temp.:
Power supply:
Lining: Electrodes: Exciter
frequency:
C
zero
: C
Span
:
System zero point:
Parameters
Setting range
Prog. Protection code:
................
0-9999 (0 = factory setting)
Language:
................
e.g., German, English, French, etc.
Flowmeter sensor
................
see name plate or the submenu "System Data / Flowmeter Primary"
Nominal size:
................
DN 1 ... DN 1000
Q
max
: ................
0.05
Q
max
DN ... 1 Q
max
DN
Pulse factor:
................
Pulse / phys. unit
Pulse width:
................
0,100 ... 2000 ms
Low cut-off setting:
................
0 ... 10% of flow range end value
Damping:
................
0.2 (0.07) ... 20 seconds
Noise reduction
................
OFF / 1 / 2/ 3 / 4
Density: ................
0.01
g/cm
3
... 5.0 g/cm
3
Unit Q
max
.:
................
e.g., l/s, l/min, l/h, hl/s, hl/min, hl/h, etc.
Unit totalizer:
................
e.g., l, hl, m
3
, igal, gal, etc.
Max. alarm:
................
%
Min. alarm:
................
%
Contact output:
................
Max. alarm, Min. alarm, Max./Min. alarm, General alarm, Empty pipe, etc.
Contact input:
................
External zero return, Totalizer reset, External system zero, no function
Current output:
................
0/4 ... 20 mA, 0/2 ... 10 mA, 0 ... 5 mA, 0 ... 10-20 mA, 4 ... 12 ... 20 mA
I
out
with alarm:
................
0 %, 103 %, 3.8 mA, Low, High
Detector e. pipe:
................
ON / OFF
Calibrate e. pipe:
................
0 ... 10000
Alarm e. pipe:
................
ON / OFF
I
out
with empty pipe:
................
0 %, 103 %, 3.8 mA, Low, High
Totalizer function:
................
Standard, difference totalizer
Display line 1:
................
Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 2:
................
Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 3:
................
Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 4:
................
Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Operating mode:
................
Standard / Piston Pump / Fast
Flow direction:
................
Supply/Return, forward
Directional display:
................
Standard, inverse
Pulse output:
Active
Passive
Communication:
HART
PROFIBUS PA
FOUNDATION Fieldbus
Without
Value E/W
Min.
Max.
Value E/W
Min.
Max.
Coil current AC
Coil voltage DC
Coil
resistance
Coil
temperature
Coil insulation resistance
DAC value
Electrode
voltage
Electrode
balance
Diagnostics
Min-Max threshold
monitoring
Error (E)
Warning (W)
Signal-to-noise
ratio
Magnetic
linearity