Initial start-up
000000-2227-967-GA-GB-281021
31
Connection to ZEISS FORUM
•
Connect the device to your network. To do this, plug the network cable
(
8
, Fig. 1) into the relevant socket (
8
•
Ensure that VISUREF 150 is connected to ZEISS FORUM and that the
following configuration parameters are set (see Table 1).
Parameter
Value
AE title of VISUREF 150
Device ID
(see
Device settings/Service/Device information
page 87)
Modality
AR
Table 1
Configuration parameters in ZEISS FORUM
For details on device connectivity to ZEISS FORUM see the FORUM
user manual and the corresponding instructions.
•
Switch to device settings.
•
Enable the interface in the
Interface/Interface enablement
menu (see
, page
•
Open the
Interface/LAN/Network socket menu (
see
, page
•
Enter the IP address and port of your FORUM server as peer address.
–
Use the format
–
Depending on the network environment, instead of the IP addresses,
the host names may be entered.
–
If no port number is defined, the FORUM standard 11119 will be used.
•
Select "DICOM" as output format and output protocol (see Fig. 11).
•
Exit the device settings and tap
Apply
.
•
VISUREF 150 is now enabled to exchange data with FORUM.