Operation of the device
000000-2227-967-GA-GB-281021
78
Fig. 50
Positioning the test eye with attached contact lens in front of the measuring head
–
Using the height adjustment buttons (
Fig. 3), adjust the motorized
chin rest so that the contact lens is set to the height of the Canthus
marks (
6
Performing contact lens measurement
•
Tap the
Contact lens
button in the upper navigation bar to enter the
contact lens measurement mode (
1
•
Select the measurement mode (
Auto
or
Manual
) (
2
, Fig. 51) to perform
measurement. Tap on the corresponding side of the two-function
button. The activated measurement mode will be displayed in white on a
blue background.
–
In
Manual
measurement mode measurement is triggered by pressing
the measurement button on the joystick.
–
In automatic mode (
Auto
) the device automatically releases a
measurement as soon as the camera image is correctly focused and
centered.
•
Now select the laterality of the contact lens to be measured (
3
Tap on
OD
(right eye) or
OS
(left eye). The activated side will be
displayed in white on a blue background.
•
In contact lens mode, the selection of illumination parameters
(
7
, Fig. 51) and the parameter control panel (
6
, Fig. 51) are disabled.