Daily startup
000000-2227-967-GA-GB-281021
35
Daily startup
WARNING - GENERAL HAZARD
Prior to using the device, the user must ensure that it is in a good
condition and fully functioning. Furthermore, the user must follow the
instructions in the user manual.
The following inspections must be carried out each working day prior to
use:
•
Visual inspection of the housing, exterior markings, user manual,
accessories and power cable to ensure that they are present and
intact. If parts are missing or damage is visible, the device should not
be used and should be taken out of service.
WARNING - RISK OF ELECTRIC SHOCK
Please take care that the following operational requirements are met
before using the device each time:
•
Use the power cable supplied with the device. The device will be
powered through the instrument table.
•
The plug must be inserted into a power outlet that has an intact
protective ground connection.
•
All cables and plugs may be used only if they are in perfect working
condition.
WARNING - RISK OF ELECTRIC SHOCK
Do not touch one or more parts shown in Fig. 5 and the patient
simultaneously. There is a risk of electric shock from current leakage.