Operation of the device
000000-2227-967-GA-GB-281021
54
1, 3, 5, and 7 measurements per eye can be selected both in
automatic and manual mode. In automatic mode, multiple
measurements are performed consecutively as soon as the eye is
focused and centered correctly. During this sequence the display is
locked for the user.
The sphere shift can be set in steps of 0.125 D and a range of -2.0 D
to +2.0 D.
Setting the illumination parameters
The device permits the adjustment of various illumination parameters for
refraction measurement. This can be helpful, particularly in the case of eyes
with clouded lenses or implanted IOL. When measuring eyes without
pathological finding, the
Standard
mode illumination parameters can be
retained.
Table 3 summarizes the illumination parameters of the device.
Illumination
parameters
Description
Scale
Reference
•
Controls the illuminance of
reference illumination of the eye
•
This parameter can be changed to
make the eye appear lighter or
darker in the live image.
from 0 (low)
to 100 (high)
in increments of 1
Ambient
•
Controls the illuminance of
ambient illumination of the eye
•
This parameter can be changed to
make the surroundings of the eye
appear lighter or darker in the live
image.
from 0 (low)
to 100 (high)
in increments of 1
Target
•
Controls the illuminance of the
fixation target in the internal
display of the device
•
Changes in this parameter may
not be perceived by the user.
from 0 (low)
to 100 (high)
in increments of 1
Table 3
Description of illumination parameters