Installation
000000-2227-967-GA-GB-281021
23
WARNING - RISK OF ELECTRIC SHOCK
If a non-medical grade LCD Monitor or PC is connected to the
autorefractor/keratometer system, the autorefractor/keratometer system
must not contact the patient. Relevant standards for non-medical
electrical equipment may require chassis leakage current limits which are
higher than that required by medical standards.
These higher limits are only acceptable outside the patient environment. It
is essential to reduce chassis leakage currents when using non-medical
electrical equipment within the patient environment. Measures for
reducing leakage current include using a medically approved isolation
transformer.
If a non-medical grade LCD monitor or PC is to be used with the
autorefractor/keratometer system, then you must ensure appropriate
mitigation to meet the safety requirements. It is the responsibility of the
user to ensure that the autorefractor/keratometer system complies with
safety requirements when combined with non-medical grade LCD
monitors or PCs.
Always use a medically-approved isolation transformer when using a non-
medical grade LCD monitor or PC.
CAUTION - PROPERTY DAMAGE
When connecting an external monitor, ensure that the monitor supports a
resolution of 800x480. Otherwise, the monitor may not display a suitable
image. If required, contact the manufacturer of the monitor.