Operation of the device
000000-2227-967-GA-GB-281021
55
The device provides the three illumination modes
Standard
,
IOL
and
User-
defined
, These modes coordinate the illumination parameters
Reference
,
Ambient
and
Target
.
•
According to the clinical application you should decide which
illumination mode is most appropriate for your measurement (see
•
If you do not change the illumination parameters, the device uses the
settings of the
Standard
illumination mode (see Table 4).
•
In case of a user-defined change of illumination parameters, the device
cannot perform keratometry. Keratometry values can only be measured
in
Standard
and
IOL
illumination mode.
Illumination mode
Application
Keratometry
Standard
Eyes without pathological
finding
yes
IOL
Eyes with implanted IOL or
eyes with slight lens clouding
(cataract)
yes
User defined
Depending on user decision
no
Table 4
Overview of illumination parameters in different illumination modes
Tap the
Illumination
button (Fig. 28) to configure the illumination
parameters. A pop-up window will open where you can set each parameter
to your specific needs.
Fig. 28
Pop-up window with unfolded illumination parameters
Summary of Contents for VISUREF 150