34
Safety
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
•
Pull out the lamp module as far as it will go.
•
Turn knob (1) through 180° until it snaps in place. This moves the
backup lamp into the illumination beam path.
•
Push the lamp module all the way back into the lamp housing.
•
Reset service hour counter (6) to "0". Use a pointed object and press
it into the recess of reset button (5).
2
Indicator: backup lamp is in use
When the red segment in knob (1) lights up, the backup bulb is in use.
Note:
If the first bulb has failed and the backup bulb is in use, make sure to
have a backup lamp module at hand as a precaution.
3
Yellow indicator lamp
Lights when the bulb has failed, or if the lamp module is defective.
After activation and ignition of the backup bulb, the yellow indicator
lamp turns off again.
4
Manual function
When the manual function has been activated, all electrical control
systems are disabled. The bulb brightness is automatically adjusted to
a fixed setting.