Technical data
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Electrical data of S8 ceiling mount
Xenon illumination system
Rated voltage
230 VAC (220...240 VAC± 10%)
115 VAC (100...125 VAC± 10%)
Current
consumption
230 VAC 3 A
120 VAC 5 A
100 VAC 6 A
Rated frequency
50...60 Hz
Fuses
Automatic circuit breaker
Electrical
outlets
– X-Y coupling
– Surgical microscope
– Remote control socket for an external signal
of a maximum of 24 V / 0.5 A.
Electrical
standard
Complying with IEC 60601-1/ EN 60601-1;
UL 60601-1; CAN/CSA-C22.2 No. 601.1-M90
Protection class I, degree of protection IPXO
Type B equipment
Product classification I
as per directive 93/42/EEC, Annex IX
Approval
EMC requirements
The system meets the EMC requirements of
IEC 60601-1-2.
The system meets the RFI requirements of Class
A (hospital grade).
The system has been designed for continuous operation.
Fiber optic illumina-
tion
Xenon short-arc reflector lamp
Color temperature: approx. 5000 K
Rated power: approx. 100 W
Backup lamp in lamp housing, manually se-
lectable.
1
2
3
1
2
3
Remote socket
View of connector side
176164