Preparations
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Attaching the equipment
Mounting the surgical microscope
Warning!
The maximum weight of the microscope including accessories must not
exceed 20 kg!
•
Bring the suspension arm into its horizontal position, pull out locking
knob (3) and turn it clockwise or counterclockwise through 180°. At the
same time, slightly move the suspension arm up and down until the
lock snaps in. This prevents the suspension arm from uncontrollably
moving upward when insufficient weight is attached.
•
Turn off the system at the power switch.
•
Hook coupling (2) with the attached surgical microscope at a slight an-
gle from above into receptacle (1) on the suspension arm, and tilt the
coupling downward into its vertical position.
•
Tighten securing screw (6) firmly using a hex key.
•
Insert cable clip (4) into opening (5) of the coupling.
•
Perform the balance setting procedure.
Warning!
Before every use and after re-equipping the system, make sure that se-
curing screw (6) has been firmly tightened!