Positioning the S8 ceiling mount
1
Working position
2
Standby position
Working position
•
Pull the ceiling mount into the working position using the handle (4).
The recommended height is approx. 1750 mm measured from the
handle to the floor.
•
When you release the handle (4), the lift arm (3) is locked in the work-
ing position (1).
Standby position
•
Press the release key of the magnetic brakes on the surgical micro-
scope. Move the suspension arm (5) to the highest possible position.
•
Push the ceiling mount upward into the parking position using the han-
dle (4). The standby position is the highest possible position.
•
When you release the handle (4), the lift arm (3) is locked in the stand-
by position (2).