Preparations
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Connecting the S88 floor stand
•
Check the voltage indicated at (3).
Caution:
The suspension system is set at the factory to the rated voltage used in
the country of destination. The rated voltage indicated at window (3) must
correspond to the rated voltage available on the site of installation. If this
is not the case, you must re-adjust the sliding switch using a suitable tool.
Note:
Insert or remove connectors (2) and (4) only if power switch (5) is off.
•
Plug the connector of the foot control panel or operating chair into con-
nector (2) of the stand.
•
Secure the microscope cable in the cable clip in such a way that it is
not stretched or bent when the microscope is turned or tilted.
•
Secure the light guide in the cable clip and insert the light guide into
the light guide socket on the microscope as far as it will go. Make sure
that the light guide is not stretched or bent when the microscope is
turned or tilted.
Use remote control socket (1) to connect external instruments with max.
24 V / 0.5 A which can be switched on / off via an AUX-signal using the
freely programmable buttons on the foot control panel.
Connect the suspension system to line power using the power cord in-
tended for it. Only use a power outlet which is provided with a properly
connected protective earth conductor.