208
Maintenance / Further information
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
SpeedFokus (autofocus) option
Zoom and focus are not set
to stored memory values
after power-on of the
system.
Power-on mode not cor-
rectly selected.
Select the desired power-on
behavior at POW ON
MODE in configuration
mode 2.
Lamp brightness changes
with magnification.
No error! Dynamic bright-
ness control has been acti-
vated.
Select the mode required at
LIGHT FUNCT in configura-
tion mode 2.
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
SpeedFokus (autofocus)
option is inoperative.
Camera control unit (CCU)
switched off.
Error code display: E20
•
Switch on MediLive
CCU.
- - -
Monitor switched off.
•
Switch on the monitor.
- - -
Failure of SpeedFokus
(autofocus) option
•
Contact service dept.
- - -
Poor focusing result or none
at all.
Focusing of the video
camera not matched to fo-
cusing of the surgical micro-
scope.
•
Focus the microscope
image, then the video
image.
- - -
Measuring field (ROI) incor-
rectly positioned.
•
Correct the ROI position
in configuration mode 4.
Suspension system moved
during focusing.
•
Do not move suspension
system during focusing.
- - -
No object in focusing range. •
Reposition surgical mi-
croscope.
- - -
No object with sufficient
contrast in focusing range.
•
Reposition surgical mi-
croscope or focus manu-
ally.
- - -