184
Operation
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Roll-over procedure:
If you have assigned FOC-MEM to key C or D of the foot control panel,
you can determine by the length of time for which you press the relevant
key of the foot control panel during operation whether a positioning run is
to be triggered (press the key for less than 2 seconds) or whether the cur-
rent position is to be saved (press the key for more than 2 seconds).
Assigning a function to the freely configurable handgrip buttons
Use the menu item "HANDGRIP:" to assign one of the following functions
to the handgrip buttons:
Roll-over procedure:
If the memory function (MEM) has been set for buttons (1) and (2) of the
handgrips, you can determine whether a positioning run is triggered or
whether the position is saved - simply by pressing the buttons for a longer
or shorter time.
If you press button (1) for less than 2 seconds, focus positioning is trig-
gered; if you press the button for more than 2 seconds, the current focus
→
AF
→
XY-RES
→
FOC-MEM
→
↑
↓
←
AUX
←
PHOTO
←
ZOOM-MEM
←
The two upper buttons (1 and 2) on the left and right handgrips of the
OPMI
Vario surgical microscope are freely configurable.
1
2
3
Display
Function of the left button (1) Function of the right button (2)
LIGHT
Reducing brightness
Increasing brightness
MEM
Focus memory
pressed for < 2 sec:
triggers positioning
Pressed for > 2 sec:
saves position
Zoom memory
pressed for < 2 sec:
triggers positioning
Pressed for > 2 sec:
saves position
PH/AUX
Camera release at the camera
interface
Triggering an AUX signal at
the AUX interface
XY-RES
XY recentering ("Reset")
XY recentering ("Reset")
AF
Triggering the autofocus option
→
XY-RES
→
AF
→
↑
↓
←
PH / AUX
←
MEM
←
LIGHT
←
MODE
STORE MANUAL
FOOT
LEFT C:
PHOTO
FOOT
RIGHT D:
PHOTO
HAND
GRIP:
LIGHT
Configuration
mode 1
SETTINGS 1