132
Preparations
G-30-1607-en
OPMI® VARIO on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 30. 01. 2009
Mounting the tube and eyepieces
•
Loosen securing screw (4) by a few turns.
•
Remove cover (3) and store it in a safe place.
•
Place binocular tube (2)
on the surgical microscope and tighten secu-
ring screw (4) firmly .
•
You can install accessory equipment between the binocular tube and
the microscope body. Lock these units in position in the same way us-
ing securing screw (4).
•
Insert widefield eyepieces (1) into the appropriate mounts as far as
they will go. The magnetic coupling reliably secures them in position.
Note:
If a documentation device is used, one of the eyepieces can be delivered
with a reticle to aid focusing. Retrofitting the reticle is only possible at the
factory or by our service representative. Always use the eyepiece with the
reticle on the same side of the binocular tube as the documentation de-
vice is installed.
Warning!
Before using and after re-equipping the unit, make sure that binocular
tube (2) is securely locked in position. Make sure that securing screw (4)
is firmly tightened!